Instrumental method for automated on-line fraction analysis and peak deconvolution in multicomponent-overloaded high-performance liquid chromatography

被引:4
作者
Cavazzini, Alberto [1 ]
Costa, Valentina [1 ]
Nadalini, Giorgio [1 ]
Dondi, Francesco [1 ]
机构
[1] Univ Ferrara, Dept Chem, I-44100 Ferrara, Italy
关键词
instrumental fraction collection; nonlinear UV-vis detector calibration; deconvolution; nonlinear and preparative chromatography; overloaded multicomponent mixtures;
D O I
10.1016/j.chroma.2006.09.090
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A bi-dimensional (non-orthogonal) chromatographic system made of two columns of different length (longer along the first direction and shorter on the second) but packed with identical (or similar) materials has been employed to perform automated on-line fraction analysis and peak deconvolution of multicomponent overloaded band profiles. Multicomponent overloaded peaks, eluted from the first column, are sampled through a remotely-controlled switching valve and fractions are injected, under analytical conditions, directly into the second direction. This set-up serves a twofold purpose. First of all, it has been possible to use linear calibration curves for detector calibration (DC). Secondly, since single component peaks were resolved on the second direction, individual concentration profiles have been obtained beneath the envelop of overlapping bands. All the information necessary to operate the bi-dimensional system can be achieved by analyzing the chromatographic behavior of the mixture on the shorter column under linear conditions. Therefore, solutions of unknown concentration can be prepared and their composition reconstructed with a simple chromatographic run. Two- and three-component overloaded mixtures have been employed to validate the system. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:36 / 41
页数:6
相关论文
共 15 条
[1]  
*AG, AG 1100 SER DIOD ARR
[2]   Weighted parallel factor analysis for calibration of HPLC-UV/Vis spectrometers in the presence of Beer's law deviations [J].
Andersson, GG ;
Dable, BK ;
Booksh, KS .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1999, 49 (02) :195-213
[3]   Combining synthetic and analytical strategies for preparative HPLC enantioseparation of monastrol racemic mixture [J].
Cavazzini, A ;
Massi, A ;
Bergamaschi, G ;
Braga, S ;
Dondi, F ;
Dondoni, A .
BIOTECHNOLOGY PROGRESS, 2004, 20 (02) :603-612
[4]   Numerical determination of the competitive isotherm of enantiomers [J].
Felinger, A ;
Cavazzini, A ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 2003, 986 (02) :207-225
[5]  
Felinger A., 1998, DATA ANAL SIGNAL PRO
[6]   The chemometric resolution and quantification of overlapped peaks form comprehensive two-dimensional liquid chromatography [J].
Fraga, CG ;
Corley, CA .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1096 (1-2) :40-49
[7]  
Giddings J.C., 1965, DYNAMICS CHROMATOGRA
[8]  
Guiochon G, 2006, Fundamentals of Preparative and Nonlinear Chromatography, V2nd
[9]   MEASUREMENT OF ADSORPTION-ISOTHERMS BY LIQUID-CHROMATOGRAPHY [J].
JACOBSON, J ;
FRENZ, J ;
HORVATH, C .
JOURNAL OF CHROMATOGRAPHY, 1984, 316 (DEC) :53-68
[10]  
KIM H, CHEM ENG SCI, V4, P1122