The spectral response of superconducting tunnel junction X-ray detectors

被引:8
作者
Friedrich, S
Fliegauf, R
Frank, M
Veldkamp, M
Labov, S
Beckhoff, B
Ulm, G
机构
[1] Lawrence Livermore Natl Lab, Adv Detector Grp, Livermore, CA 94550 USA
[2] Phys Tech Bundesanstalt, D-10587 Berlin, Germany
关键词
superconducting tunnel junctions; X-ray spectrometers; STJ line shapes; STJ response functions; synchrotron detectors; superconducting devices; ENERGY RESOLUTION; PERFORMANCE;
D O I
10.1016/j.nima.2005.07.041
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The Advanced Detector Group at LLNL is developing superconducting Nb-Al-AlOx-Al-Nb tunnel junction detector arrays for high-resolution soft X-ray spectrometry. We have characterized the detector response function in the energy range between 50 and 2000 eV with monochromatic X-rays, in collaboration with the Physikalisch-Technische Bundesanstalt Laboratory at the BESSY II electron storage ring. The detectors have an energy resolution between 7 and 15 eV FWHM in that energy range, and a count rate capability of similar to 10,000 counts/s per detector pixel. In addition, the detector response function exhibits several spectral artifacts, and although their magnitude is only on the order of 1 % of the main line, they can reduce the sensitivity for X-ray analysis of dilute specimens if they overlap with the correspondingly weak emission lines. We have characterized these artifacts by scanning a collimated synchrotron radiation beam across the detector area. We identify their origin due to X-ray absorption at locations other than the main Nb absorber film, and outline future detector designs to reduce the artifacts and thereby extend the spectrometer sensitivity. (c) 2005 Published by Elsevier B.V.
引用
收藏
页码:35 / 45
页数:11
相关论文
共 28 条
[1]   Energy resolution of 12 eV at 5.9 keV from Al-superconducting tunnel junction detectors [J].
Angloher, G ;
Hettl, P ;
Huber, M ;
Jochum, J ;
von Feilitzsch, F ;
Mössbauer, RL .
JOURNAL OF APPLIED PHYSICS, 2001, 89 (02) :1425-1429
[2]   A SIMPLE AND ROBUST NIOBIUM JOSEPHSON JUNCTION INTEGRATED-CIRCUIT PROCESS [J].
BARFKNECHT, AT ;
RUBY, RC ;
KO, H .
IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) :3125-3128
[3]   Characterization of an Nb/Al/AlOx/Al/Nb superconducting tunnel junction detector with a very high spatial resolution in the soft X-ray range [J].
Bechstein, S ;
Beckhoff, B ;
Fliegauf, R ;
Weser, J ;
Ulm, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2004, 59 (02) :215-221
[4]   TXRF analysis of low Z elements and TXRF-NEXAFS speciation of organic contaminants on silicon wafer surfaces excited by monochromatized undulator radiation [J].
Beckhoff, B ;
Fliegauf, R ;
Ulm, G ;
Weser, J ;
Pepponi, G ;
Streli, C ;
Wobrauschek, P ;
Ehmann, T ;
Fabry, L ;
Pahlke, S ;
Kanngiesser, B ;
Malzer, W .
ULTRA CLEAN PROCESSING OF SILICON SURFACES V, 2003, 92 :165-169
[5]   Investigation of high-resolution superconducting tunnel junction detectors for low-energy X-ray fluorescence analysis [J].
Beckhoff, B ;
Fliegauf, R ;
Ulm, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2003, 58 (04) :615-626
[6]   X-ray detector calibration in the PTB radiometry laboratory at the electron storage ring BESSY II [J].
Beckhoff, B ;
Klein, R ;
Krumrey, M ;
Scholze, F ;
Thornagel, R ;
Ulm, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 444 (1-2) :480-483
[7]   Metallic magnetic calorimeters (MMC): detectors for high-resolution X-ray spectroscopy [J].
Fleischmann, A ;
Link, M ;
Daniyarov, T ;
Rotzinger, H ;
Enss, C ;
Seidel, GM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 520 (1-3) :27-31
[8]   Energy resolution and high count rate performance of superconducting tunnel junction x-ray spectrometers [J].
Frank, M ;
Hiller, LJ ;
le Grand, JB ;
Mears, CA ;
Labov, SE ;
Lindeman, MA ;
Netel, H ;
Chow, D ;
Barfknecht, AT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (01) :25-31
[9]   A multichannel superconducting tunnel junction detector for high-resolution X-ray spectroscopy of magnesium diboride films [J].
Friedrich, S ;
Vailionis, A ;
Drury, O ;
Niedermayr, T ;
Funk, T ;
Kang, WN ;
Choi, EM ;
Kim, HJ ;
Lee, SI ;
Cramer, SP ;
Kim, C ;
Labov, SE .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2003, 13 (02) :1114-1119
[10]   Line-splitting in high-resolution superconducting tunnel junction EUV detectors [J].
Friedrich, S ;
Hiller, LJ ;
Cunningham, MF ;
Labov, SE .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (01) :836-839