Survival rate, fracture strength and failure mode of ceramic implant abutments after chewing simulation

被引:106
作者
Butz, F
Heydecke, G
Okutan, M
Strub, JR
机构
[1] Univ Hosp, Sch Dent, Dept Prosthodont, D-79106 Freiburg, Germany
[2] Univ Calif Los Angeles, Sch Dent, Div Adv Prosthodont Biomat & Hosp Dent, Jane & Jerry Weintraub Ctr Reconstruct Biotechnol, Los Angeles, CA 90024 USA
关键词
dental implants; ceramic implant abutment; fracture strength; failure mode; survival rate; dynamic loading;
D O I
10.1111/j.1365-2842.2005.01515.x
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
The aim of this study was to compare titanium-reinforced ZrO2 and pure Al2O3 abutments regarding their outcome after chewing simulation and static loading. Forty-eight standard diameter implants with an external hexagon were divided into three groups of 16 implants each and restored with three different types of abutments (group A: ZrO2 abutments with titanium inserts; group B: Al2O3 abutments; group C: titanium abutments). All abutments were fixated on the implants with gold-alloy screws at 32 Ncm torque, and metal crowns were adhesively cemented onto the abutments. The specimens were exposed to 1.2 million cycles in a chewing simulator. Surviving specimens were subsequently loaded until fracture in a static testing device. Fracture loads (N) and fracture modes were recorded. A Wilcoxon Rank test to compare fracture loads among the three groups and a Fisher exact test to detect group differences in fracture modes were used for statistical evaluation (P < 0.05). All specimens but one of group B survived chewing simulation. No screw loosening occurred. The median fracture loads (+/- s.d.) were as follows: group A, 294 N (+/- 53); group B, 239 N (+/- 83), and group C, 324 N (+/- 85). The smaller fracture loads in group B were statistically significant. The use of pure Al2O3 abutments resulted in significantly more abutment fractures. It is proposed that titanium-reinforced ZrO2 abutments perform similar to metal abutments, and can therefore be recommended as an aesthetic alternative for the restoration of single implants in the anterior region. All-ceramic abutments made of Al2O3 possess less favourable properties.
引用
收藏
页码:838 / 843
页数:6
相关论文
共 37 条
[1]
The mucosal attachment at different abutments - An experimental study in dogs [J].
Abrahamsson, I ;
Berglundh, T ;
Glantz, PO ;
Lindhe, J .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1998, 25 (09) :721-727
[2]
Andersson B, 2003, INT J PROSTHODONT, V16, P640
[3]
Andersson B, 2001, INT J PROSTHODONT, V14, P432
[4]
Bates J F, 1975, J Oral Rehabil, V2, P281, DOI 10.1111/j.1365-2842.1975.tb00921.x
[5]
Binon PP, 2000, INT J ORAL MAX IMPL, V15, P76
[6]
Binon PP, 2000, QUINTESSENCE DENT TE, V23, P91
[7]
BRODBECK U, 2003, J ESTHET RESTOR DENT, V15, P24
[8]
Brodbeck Urs, 2003, J Esthet Restor Dent, V15, P10, DOI 10.1111/j.1708-8240.2003.tb00278.x
[9]
Examination of the implant-abutment interface after fatigue testing [J].
Cibirka, RM ;
Nelson, SK ;
Lang, BR ;
Rueggeberg, FA .
JOURNAL OF PROSTHETIC DENTISTRY, 2001, 85 (03) :268-275
[10]
FUNCTIONAL OCCLUSAL FORCES - INVESTIGATION BY TELEMETRY [J].
DEBOEVER, JA ;
MCCALL, WD ;
HOLDEN, S ;
ASH, MM .
JOURNAL OF PROSTHETIC DENTISTRY, 1978, 40 (03) :326-333