New access device for hemodialysis

被引:14
作者
Levin, NW
Yang, PM
Hatch, DA
Dubrow, AJ
Caraiani, NS
Ing, TS
Gandhi, VC
Alto, A
Davila, SM
Prosl, FR
Polaschegg, HD
Megerman, J
机构
[1] Biolink Corp, Clin Affairs, Middleboro, MA 02346 USA
[2] Renal Res Inst, New York, NY USA
[3] Beth Israel Deaconess Med Ctr, New York, NY 10003 USA
[4] US Dept Vet Affairs, Vet Affairs Edward Hines Jr Hosp, Hines, IL 60141 USA
关键词
D O I
10.1097/00002480-199809000-00042
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new subcutaneous device (Dialock(TM); Biolink Corp., Middleboro, MA) provides vascular access to patients who currently require hemodialysis (HD). The device consists of a port-like valve, implanted subcutaneously below the clavicle, which provides a linear flow passage to two catheters placed in the right atrium via the jugular vein. The valve is accessed percutaneously with needle-cannulas that functionally convert the device to twin catheters for connecting the patient to the HD lines. Interdialytic patency is maintained using a standard heparin lock. The device has been implanted in 10 outpatients under local anesthesia, with almost immediate use for HD (median, 3 days) and has functioned successfully for more than 6 months (mean +/- SD, 4.0 +/- 1.7; > 400 dialysis sessions). Blood flows over 300 ml/min were consistently achieved (average, 320 +/- 50) with venous and arterial pressures of 197 +/- 42 mmHg and -241 +/- 31 mmHg, respectively. After 40 patient-months, condition of the needle puncture sites remains satisfactory. Four systemic infections have occurred in three patients; all have resolved without the need for device removal. There have been no infections at the puncture sites. One patient whose heparin lock was not changed for 23 days (for reasons unrelated to the device) required fibrin sheath stripping of his catheters. Patient and nurse acceptance has been excellent. The device may offer substantial improvement over conventional devices for HD access.
引用
收藏
页码:M529 / M531
页数:3
相关论文
共 12 条
[11]  
*US REN DAT SYST, 1997, USRDS 1997 ANN DAT R
[12]   VANCOMYCIN STABILITY IN HEPARIN AND TOTAL PARENTERAL-NUTRITION SOLUTIONS - NOVEL-APPROACH TO THERAPY OF CENTRAL VENOUS CATHETER-RELATED INFECTIONS [J].
YAO, JDC ;
ARKIN, CF ;
KARCHMER, AW .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1992, 16 (03) :268-274