Fully automated immunoassay system of endocrine disrupting chemicals using monoclonal antibodies chemically conjugated to bacterial magnetic particles

被引:50
作者
Matsunaga, T
Ueki, F
Obata, K
Tajima, H
Tanaka, T
Takeyama, H
Goda, Y
Fujimoto, S
机构
[1] Tokyo Univ Agr & Technol, Dept Biotechnol, Fac Technol, Koganei, Tokyo 1848588, Japan
[2] Precis Syst Sci Co Ltd, Matsudo, Chiba 2700064, Japan
[3] Takeda Chem Ind Ltd, Osaka 5328686, Japan
关键词
bacterial magnetic particles (BMPs); endocrine disrupting chemicals (EDCs); fully automated immunoassay system; alkylphenol; ethoxylates (APES); bisphenol A (BPA); linear alkylbenzene sulfonates (LASs);
D O I
10.1016/S0003-2670(02)01036-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The development of a rapid and high-throughput detection system for endocrine disrupting chemicals (EDCs) has been required in the recent years. A fully automated immunoassay system was described for the detection of EDCs, such as alkylphenol ethoxylates (APEs), bisphenol A (BPA) and linear alkylbenzene sulfonates (LASS), using monoclonal antibodies chemically conjugated to bacterial magnetic particles (BMPs) and alkaline phosphatase (ALP)-conjugated EDCs. EDC concentrations were evaluated by the decrease in luminescence based on the competitive reaction of EDCs and ALP-conjugated EDCs. Full automation of the BMP-based immunoassay was achieved by using an automated eight-way pipet moving at X-, Y- and Z-direction and a B/F separation unit. B/F separation was performed on the tip surface of eight-way pipet with a retractable magnet mounted close to the pipet tip. Immunoreactions were saturated after 10 min, and the assay was completed within 15 min. The detection ranges for APE, BPA and LAS were 6.6 ppb-66 ppm, 2.3 ppt-2.3 ppm, and 35 ppt-35 ppm, respectively. This BMP-based immunoassay system has advantages due to its high sensitivity and rapid measurement of samples. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:75 / 83
页数:9
相关论文
共 28 条
[1]   MAGNETOTACTIC BACTERIA [J].
BLAKEMORE, R .
SCIENCE, 1975, 190 (4212) :377-379
[2]   Direct competitive ELISA for the determination of polychlorinated biphenyls in soil samples [J].
Deng, AP ;
Kolár, V ;
Ulrich, R ;
Fránek, M .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 373 (08) :685-690
[3]   Analysis of nonylphenol polyethoxylates and their degradation products in river water and sewage effluent by gas chromatography ion trap (tandem) mass spectrometry with electron impact and chemical ionization [J].
Ding, WH ;
Tzing, SH .
JOURNAL OF CHROMATOGRAPHY A, 1998, 824 (01) :79-90
[4]   Analysis of linear alkylbenzenesulfonates in water samples by large-volume injection-port derivatization and gas chromatography-mass spectrometry [J].
Ding, WH ;
Chen, CT .
JOURNAL OF CHROMATOGRAPHY A, 1999, 857 (1-2) :359-364
[5]   An enzyme-linked immunosorbent assay for detection of linear alkylbenzene sulfonate: Development and field studies [J].
Fujita, M ;
Ike, M ;
Goda, Y ;
Fujimoto, S ;
Toyoda, Y ;
Miyagawa, KI .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1998, 32 (08) :1143-1146
[6]   Development of the ELISAs for detection of hormone-disrupting chemicals [J].
Goda, Y ;
Kobayashi, A ;
Fukuda, K ;
Fujimoto, S ;
Ike, M ;
Fujita, M .
WATER SCIENCE AND TECHNOLOGY, 2000, 42 (7-8) :81-88
[7]   MAGNETIC BEAD ANTIGEN CAPTURE ENZYME-LINKED IMMUNOASSAY IN MICROTITRE TRAYS FOR RAPID DETECTION OF SCHISTOSOMAL CIRCULATING ANODIC ANTIGEN [J].
GUNDERSEN, SG ;
HAAGENSEN, I ;
JONASSEN, TO ;
FIGENSCHAU, KJ ;
DEJONGE, N ;
DEELDER, AM .
JOURNAL OF IMMUNOLOGICAL METHODS, 1992, 148 (1-2) :1-8
[8]   TOXIC EFFECTS OF LINEAR ALKYLBENZENE SULFONATE ON THE TIGER PRAWN, PENAEUS-MONODON [J].
HWANG, DF ;
CHEN, MY ;
YOSHIDA, T ;
JENG, SS .
ECOTOXICOLOGY AND ENVIRONMENTAL SAFETY, 1993, 26 (03) :285-292
[9]  
INOUE K, 2002, B ANAL TECHNOL BIOME, V773, P97
[10]   Nonylphenol ethoxylates, but not nonylphenol, are substrates of the human multidrug resistance P-glycoprotein [J].
Loo, TW ;
Clarke, DM .
BIOCHEMICAL AND BIOPHYSICAL RESEARCH COMMUNICATIONS, 1998, 247 (02) :478-480