Capillary electrophoresis-dynamic reaction cell inductively coupled plasma mass spectrometry (CE-DRC-ICP-MS) for the speciation of iron(III/II), vanadium(V/IV) and chromium(VI/III) is described. Two different CE migration modes were employed for separating the six metal ions using pre-capillary complexation. One is counter-electroosmotic mode in which iron(III/II) and vanadium(V/IV) ions were well separated using a 60 cm x 75 mum i.d. fused silica capillary. The voltage was set at +22 kV and a 15 mmol l(-1) tris(hydroxymethyl)aminomethane (Tris) buffer (pH 8.75) containing 0.5 mmol l(-1) ethylenediaminetetraacetic acid (EDTA) and 0.5 mmol l(-1) ortho-phenanthroline (phen) was used as the electrophoretic buffer. The other is co-electroosmotic mode in which chromium(VI/III) ions were well separated while the applied voltage was set at -22 kV and a 10 mmol l(-1) ammonium citrate buffer (pH 7.7) containing 0.5 mmol l(-1) diethylenetriaminepentaacetic acid (DTPA) and 0.01% polybrene was used as the electrophoretic buffer. The mass spectra were measured at m/z 51, 52 and 56 for V, Cr and Fe, respectively. The interfering polyatomic ions of (ClO+)-Cl-35-O-16, (ArC+)-Ar-40-C-12 and (ArO+)-Ar-40-O-16 on V-51(+), Cr-52(+) and Fe-56(+) determination were reduced in intensity significantly by using NH3 as the reaction cell gas in the DRC. The detection limits were in the range of 0.1-0.5, 0.4-1.3 and 1.2-1.7 ng ml(-1) for V, Cr and Fe, respectively. Applications of the method for the speciation of V, Cr and Fe in wastewater were demonstrated. The recoveries were in the range of 92-120% for various species. (C) 2004 Elsevier B.V. All rights reserved.