Optimization of electrode alignment for electrochemical detection in capillary electrophoresis using a scanning electrochemical microscope

被引:9
作者
Roach, DM [1 ]
Hooper, SE [1 ]
Anderson, MR [1 ]
机构
[1] Virginia Polytech Inst & State Univ, Dept Chem, Blacksburg, VA 24061 USA
关键词
electrophoresis; scanning electrochemical microscopy; electrochemical detection; electrode alignment; detection optimization;
D O I
10.1002/elan.200503368
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Electrochemical detection in capillary electrophoresis requires decoupling the voltage applied to the working electrode from the separation voltage applied across the capillary. End-capillary electrochemical detection achieves this by placing the electrode just outside the ground end of the separation capillary. Obtaining adequate signal-to-noise in this arrangement requires using small inner diameter capillaries. Decreasing the inner diameter of the separation capillary, however, increases the difficulty of aligning the microelectrode with the open end of the capillary. Using scanning electrochemical microscopy (SECM), the position of the capillary opening is determined while electroactive material is continuously emerging from the end of the capillary. The SECM instrument is then used to place the electrode at the position of maximum current for subsequent separations. Subsequent measurements found that the best signal-to-noise is obtained when the detection electrode is placed directly opposite the capillary opening and just outside of the capillary opening. When the electrode is further above the opening (but still opposite the capillary opening), the signal-to-noise does not dramatically decrease until the electrode is more than 30 mu m above the 10 mu m inner-diameter capillary.
引用
收藏
页码:2254 / 2259
页数:6
相关论文
共 23 条
[1]   Determination of monoamine transmitters and tyrosine in biological samples by capillary electrophoresis with electrochemical detection [J].
Chen, G ;
Ye, JN ;
Cheng, JS .
CHROMATOGRAPHIA, 2000, 52 (3-4) :137-141
[2]   AN ELECTROCHEMICAL COIL FOR END-COLUMN AMPEROMETRIC DETECTION IN CAPILLARY ELECTROPHORESIS [J].
CHEN, MC ;
HUANG, HJ .
ANALYTICAL CHEMISTRY, 1995, 67 (21) :4010-4014
[3]   Rudimentary capillary-electrode alignment for capillary electrophoresis with electrochemical detection [J].
Fermier, AM ;
Gostkowski, ML ;
Colon, LA .
ANALYTICAL CHEMISTRY, 1996, 68 (09) :1661-1664
[4]  
Holland LA, 1998, ANAL COMMUN, V35, p1H
[5]   END-COLUMN DETECTION FOR CAPILLARY ZONE ELECTROPHORESIS [J].
HUANG, XH ;
ZARE, RN ;
SLOSS, S ;
EWING, AG .
ANALYTICAL CHEMISTRY, 1991, 63 (02) :189-192
[6]   MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY OF PUNGENT COMPOUNDS USING SIMULTANEOUS ONLINE ULTRAVIOLET AND ELECTROCHEMICAL DETECTION [J].
KHALED, MY ;
ANDERSON, MR ;
MCNAIR, HM .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1993, 31 (07) :259-264
[7]   END-COLUMN ELECTROCHEMICAL DETECTION FOR INORGANIC AND ORGANIC-SPECIES IN HIGH-VOLTAGE CAPILLARY ELECTROPHORESIS [J].
LU, WZ ;
CASSIDY, RM ;
BARANSKI, AS .
JOURNAL OF CHROMATOGRAPHY, 1993, 640 (1-2) :433-440
[8]   ELECTROCHEMICAL DETECTION WITH MICROELECTRODES IN CAPILLARY-FLOW SYSTEMS [J].
MATYSIK, FM ;
MEISTER, A ;
WERNER, G .
ANALYTICA CHIMICA ACTA, 1995, 305 (1-3) :114-120
[9]   Experimental characterization of end-column electrochemical detection in conjunction with nonaqueous capillary electrophoresis [J].
Matysik, FM .
ANALYTICAL CHEMISTRY, 2000, 72 (11) :2581-2586
[10]  
Matysik FM, 2000, ELECTROANAL, V12, P1349, DOI 10.1002/1521-4109(200011)12:17<1349::AID-ELAN1349>3.3.CO