Sevoflurane degradation to compound A in anaesthesia breathing systems

被引:29
作者
Cunningham, DD
Huang, S
Webster, J
Mayoral, J
Grabenkort, RW
机构
[1] Abbott Laboratories, Hosp. Prod. Div. R. and D., Department 971, Abbott Park, IL 60064-3500, One Abbott Park Road
关键词
anaesthetics volatile; sevoflurane; carbon dioxide; absorption; equipment; breathing systems;
D O I
10.1093/bja/77.4.537
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Determination of an effective rate constant and activation energy allowed the application of steady-state theory to predict concentrations of compound A from sevoflurane concentrations, fresh gas flow rate, absorbent temperature and amount of absorbent. Studies by eight research groups were compared. Lower concentrations of compound A than predicted were observed at low flow rates, suggesting that its degradation by the absorbent is important in limiting the maximum observed concentrations in closed and low-flow breathing systems. Trial-to-trial and batch-to-batch variations in compound A concentrations were observed in model system tests of commercial and pilot-plant absorbents. Chemical modification of the absorbent with glycerol lowered concentrations of compound A, possibly by formation of a nucleophilic addition product. An ideal chemical scavenger would remain stable and non-volatile in the absorbent before irreversibly reacting with compound A to form a stable non-volatile product.
引用
收藏
页码:537 / 543
页数:7
相关论文
共 22 条
[1]   CONCERNING THE FORMATION OF HYDROGEN IN NUCLEAR WASTE - QUANTITATIVE GENERATION OF HYDROGEN VIA A CANNIZZARO INTERMEDIATE [J].
ASHBY, EC ;
DOCTOROVICH, F ;
LIOTTA, CL ;
NEUMANN, HM ;
BAREFIELD, EK ;
KONDA, A ;
ZHANG, K ;
HURLEY, J ;
SIEMER, DD .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1993, 115 (03) :1171-1173
[2]   EFFECT OF TOTAL FLOW-RATE ON THE CONCENTRATION OF DEGRADATION PRODUCTS GENERATED BY REACTION BETWEEN SEVOFLURANE AND SODA LIME [J].
BITO, H ;
IKEDA, K .
BRITISH JOURNAL OF ANAESTHESIA, 1995, 74 (06) :667-669
[3]   LONG-DURATION, LOW-FLOW SEVOFLURANE ANESTHESIA USING 2 CARBON-DIOXIDE ABSORBENTS - QUANTIFICATION OF DEGRADATION PRODUCTS IN THE CIRCUIT [J].
BITO, H ;
IKEDA, K .
ANESTHESIOLOGY, 1994, 81 (02) :340-345
[4]   DEGRADATION PRODUCTS OF SEVOFLURANE DURING LOW-FLOW ANESTHESIA [J].
BITO, H ;
IKEDA, K .
BRITISH JOURNAL OF ANAESTHESIA, 1995, 74 (01) :56-59
[5]  
BITO H, 1994, ANESTH ANALG, V79, P946
[6]   CLOSED-CIRCUIT ANESTHESIA WITH SEVOFLURANE IN HUMANS - EFFECTS ON RENAL AND HEPATIC-FUNCTION AND CONCENTRATIONS OF BREAKDOWN PRODUCTS WITH SODA LIME IN THE CIRCUIT [J].
BITO, H ;
IKEDA, K .
ANESTHESIOLOGY, 1994, 80 (01) :71-76
[7]   PERFORMANCE OF ABSORBENTS - EFFECT OF MOISTURE [J].
BROWN, ES ;
BAKAMJIAN, V ;
SENIFF, AM .
ANESTHESIOLOGY, 1959, 20 (05) :613-617
[8]   THE ACTIVITY AND SURFACE AREA OF FRESH SODA LIME [J].
BROWN, ES .
ANESTHESIOLOGY, 1958, 19 (02) :208-212
[9]   ANALYSIS OF SEVOFLURANE DEGRADATION PRODUCTS IN VAPOR-PHASE SAMPLES [J].
CUNNINGHAM, DD ;
WEBSTER, J ;
NELSON, D ;
WILLIAMSON, B .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1995, 668 (01) :41-52
[10]  
FANG ZX, 1995, UNPUB PUBLIC COMMENT