Polyethylene cross-linking by two different methods reduces acetabular liner wear in a hip joint wear simulator

被引:42
作者
D'Lima, DD [1 ]
Hermida, JC [1 ]
Chen, PC [1 ]
Colwell, CW [1 ]
机构
[1] Scripps Clin Ctr Orthopaed Res & Educ, Joint Mech Lab, La Jolla, CA 92037 USA
关键词
D O I
10.1016/S0736-0266(02)00240-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Advances in cross-linking have led to the development of wear resistant ultrahigh molecular weight polyethylene for total joint replacement. This study compared wear reduction by two different cross-linking methods as measured in a hip wear simulator One highly cross-linked polyethylene was treated with 7.5 Mrad gamma irradiation with post-irradiation annealing and a sterilization dose of 2.5 Mrad (10 Gamma), while the other used 9.5 Mrad warm irradiation with 10 MeV, electron-beam (9.5 EB); Liners of the same design, made from nominally cross-linked (gamma sterilized) polyethylene were also tested. Gravimetric wear analysis was performed every 500,000 cycles for 5,000,000 cycle. After correcting for weight gain due to water absorption, the nominally cross linked liners demonstrated mean wear rates of 15.7 (+/-1.7) and 12.5 (+/-1.0) mg/million cycles. Both highly cross-linked polyethylene liners demonstrated significantly less wear than their respective controls (with mean rates of 1.5 (+/-1.2) and -1.4 ( +/-1.5) mg/ million cycle). The 9.5 EB liners gained Weight presumably due to increased fluid absorption, in addition, to that measured in loaded-soaked control implants. Any wear occuring was therefore assumed to have been more than offset by weight gain. Highly cross-linked polyethylene was significantly more wear resistant than non- or nominally cross-linked polyethylene. The differences in wear rates between the two highly cross-linked polyethylene designs (9.5 EB or 10 Gamma) are probably too small to be clinically significant. (C) 2003 Orthopaedic Research Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:761 / 766
页数:6
相关论文
共 49 条
[1]
AMSTUTZ HC, 1992, CLIN ORTHOP RELAT R, P7
[2]
IMPROVED CEMENTING TECHNIQUES AND FEMORAL COMPONENT LOOSENING IN YOUNG-PATIENTS WITH HIP-ARTHROPLASTY - A 12-YEAR RADIOGRAPHIC REVIEW [J].
BARRACK, RL ;
MULROY, RD ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :385-389
[3]
BLACK J, 1996, CLIN ORTHOP S, V329, P244, DOI DOI 10.1097/00003086-199608001-00021
[4]
Bono J V, 1994, J Arthroplasty, V9, P119, DOI 10.1016/0883-5403(94)90059-0
[5]
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[6]
BRAGDON CR, 1999, T 45 ORTH RES SOC, V24, P824
[7]
BRAGDON CR, 1998, T SOC BIOMATER, V21, P2
[8]
Busanelli L, 1996, Chir Organi Mov, V81, P263
[9]
Callaway GH, 1995, J ARTHROPLASTY, V10, P855
[10]
Early failure of Hylamer acetabular inserts due to eccentric wear [J].
Chmell, MJ ;
Poss, R ;
Thomas, WH ;
Sledge, CB .
JOURNAL OF ARTHROPLASTY, 1996, 11 (03) :351-353