Analytical Comparison of Three Quantitative Immunochemical Fecal Occult Blood Tests for Colorectal Cancer Screening

被引:36
作者
Guittet, Lydia [1 ]
Guillaume, Elodie
Levillain, Romuald [2 ]
Beley, Philippe [2 ]
Tichet, Jean [2 ]
Lantieri, Olivier [2 ]
Launoy, Guy
机构
[1] UCBN, CHU Caen, INSERM ERI3 Canc & Populat, UFR Med,Univ Hosp, F-14000 Caen, France
[2] IRSA Inst Interreg Sante, La Riche, France
关键词
COLONIC LESIONS; PERFORMANCE; NUMBER;
D O I
10.1158/1055-9965.EPI-10-0594
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The superiority of several immunochemical fecal occult blood tests (I-FOBT) over guaiac-based tests in colorectal cancer screening is now established. The aim of this study was to compare the analytical performance of 3 quantitative I-FOBTs. Methods: Stool samples from 10 healthy volunteers, initially I-FOBT negative, supplemented with human blood, were used to compare reproducibility and stability of measurement at varying storage temperatures (4 degrees C, 10 degrees C, 20 degrees C, and 30 degrees C) and durations before test analysis (1 to 10 days) for 3 I-FOBTs (New Hemtube/Magstream HT, OC-Auto sampling bottle3/OC-Sensor DIANA, and FOB Gold/SENTiFOB). Concentrations ranging from 0 to 350 mu g Hb/g of feces were evaluated. Results: The measurement reproducibility of OC-Sensor was superior to Magstream and far superior to FOB Gold. For all tests, variability was essentially related to sampling. Detected hemoglobin (Hb) levels were substantially lower for all tests at temperatures above 20 degrees C. At 20 degrees C, this loss in concentration was less important with OC-Sensor (significant 1.7% daily decrease vs. 7.4% for Magstream and 7.8% for FOB Gold). At 30 degrees C, daily loss was 8.6% with OC-Sensor, whereas after 24 hours, only 30% of the original Hb was detected with FOB Gold, compared to 70% with Magstream. No Hb was detected on day 5 for the latter 2 tests. Conclusions: About reproducibility and temperature stability, OC-Sensor performed better than Magstream and far better that FOB Gold. Impact: Independently of the chosen test, the delay between sampling and test processing should be reduced, the maximal admissible delay depending on ambient temperature. Cancer Epidemiol Biomarkers Prev; 20(7); 1492-501. (C) 2011 AACR.
引用
收藏
页码:1492 / 1501
页数:10
相关论文
共 18 条
[1]  
*AUSTR GOVT DEP HL, NAT BOW CANC SCREEN
[2]   Screening for Colorectal Cancer: US Preventive Services Task Force Recommendation Statement [J].
Calonge, Ned ;
Petitti, Diana B. ;
DeWitt, Thomas G. ;
Dietrich, Allen J. ;
Gregory, Kimberly D. ;
Harris, Russell ;
Isham, George ;
LeFevre, Michael L. ;
Leipzig, Roseanne M. ;
Loveland-Cherry, Carol ;
Marion, Lucy N. ;
Melnyk, Bernadette ;
Moyer, Virginia A. ;
Ockene, Judith K. ;
Sawaya, George F. ;
Yawn, Barbara P. .
ANNALS OF INTERNAL MEDICINE, 2008, 149 (09) :627-+
[3]   Association of FOBT-assessed faecal Hb content with colonic lesions detected in the Florence screening programme [J].
Ciatto, S. ;
Martinelli, F. ;
Castiglione, G. ;
Mantellini, P. ;
Rubeca, T. ;
Grazzini, G. ;
Bonanomi, A. G. ;
Confortini, M. ;
Zappa, M. .
BRITISH JOURNAL OF CANCER, 2007, 96 (02) :218-221
[4]   Immunochemical faecal occult blood test: number of samples and positivity cutoff. What is the best strategy for colorectal cancer screening? [J].
Grazzini, G. ;
Visioli, C. B. ;
Zorzi, M. ;
Ciatto, S. ;
Banovich, F. ;
Bonanomi, A. G. ;
Bortoli, A. ;
Castiglione, G. ;
Cazzola, L. ;
Confortini, M. ;
Mantellini, P. ;
Rubeca, T. ;
Zappa, M. .
BRITISH JOURNAL OF CANCER, 2009, 100 (02) :259-265
[5]   Influence of seasonal variations in ambient temperatures on performance of immunochemical faecal occult blood test for colorectal cancer screening: observational study from the Florence district [J].
Grazzini, Grazia ;
Ventura, Leonardo ;
Zappa, Marco ;
Ciatto, Stefano ;
Confortini, Massimo ;
Rapi, Stefano ;
Rubeca, Tiziana ;
Visioli, Carmen Beatriz ;
Halloran, Stephen P. .
GUT, 2010, 59 (11) :1511-1515
[6]   Comparison of a guaiac based and an immunochemical faecal occult blood test in screening for colorectal cancer in a general average risk population [J].
Guittet, L. ;
Bouvier, V. ;
Mariotte, N. ;
Vallee, J. P. ;
Arsene, D. ;
Boutreux, S. ;
Tichet, J. ;
Launoy, G. .
GUT, 2007, 56 (02) :210-214
[7]   Comparison of a guaiac and an immunochemical faecal occult blood test for the detection of colonic lesions according to lesion type and location [J].
Guittet, L. ;
Bouvier, V. ;
Mariotte, N. ;
Vallee, J. P. ;
Levillain, R. ;
Tichet, J. ;
Launoy, G. .
BRITISH JOURNAL OF CANCER, 2009, 100 (08) :1230-1235
[8]   Performance of immunochemical faecal occult blood test in colorectal cancer screening in average-risk population according to positivity threshold and number of samples [J].
Guittet, Lydia ;
Bouvier, Veronique ;
Mariotte, Nicole ;
Vallee, Jean-Pierre ;
Levillain, Romuald ;
Tichet, Jean ;
Launoy, Guy .
INTERNATIONAL JOURNAL OF CANCER, 2009, 125 (05) :1127-1133
[9]   Cochrane systematic review of colorectal cancer screening using the fecal occult blood test (Hemoccult): An update [J].
Hewitson, Paul ;
Glasziou, Paul ;
Watson, Eila ;
Towler, Bernie ;
Irwig, Les .
AMERICAN JOURNAL OF GASTROENTEROLOGY, 2008, 103 (06) :1541-1549
[10]   Screening for colorectal cancer: random comparison of guaiac and immunochemical faecal occult blood testing at different cut-off levels [J].
Hol, L. ;
Wilschut, J. A. ;
van Ballegooijen, M. ;
van Vuuren, A. J. ;
van der Valk, H. ;
Reijerink, J. C. I. Y. ;
van der Togt, Acm ;
Kuipers, E. J. ;
Habbema, J. D. F. ;
van Leerdam, Me .
BRITISH JOURNAL OF CANCER, 2009, 100 (07) :1103-1110