Impact of Tube Current-Time and Tube Voltage Reduction in 64-Detector-row Computed Tomography Pulmonary Angiography for Pulmonary Embolism in a Porcine Model

被引:22
作者
Henes, Frank Oliver Gerhard [1 ]
Groth, Michael [1 ]
Begemann, Philipp G. C. [1 ]
Bley, Thorsten [1 ]
Adam, Gerhard [1 ]
Regier, Marc [1 ]
机构
[1] Univ Med Ctr Hamburg Eppendorf, Dept Diagnost & Intervent Radiol, D-20246 Hamburg, Germany
关键词
computed tomography; computed tomography angiography; pulmonary angiography; pulmonary embolism; radiation dosage; dose reduction; DOSE CHEST CT; HELICAL CT; IMAGE QUALITY; RADIATION-EXPOSURE; PREGNANT PATIENTS; FOLLOW-UP; FEASIBILITY; PROTOCOLS; RESOLUTION; DIAGNOSIS;
D O I
10.1097/RTI.0b013e31820b8696
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: To evaluate the impact of dose reduction in multidetector computed tomography pulmonary angiography (CTPA) for detection of pulmonary embolism (PE). Materials and Methods: After induction of PE in 6 anesthetized pigs, a 64-detector-row CTPA was performed at 3 different dose protocols: A (120 kV/120mAs(eff).), B (120 kV/80mAs(eff).), and C (80 kV/80mAs(eff).). Images were evaluated by 2 radiologists independently. A high-dose CTPA (120 kV/250mAs(eff).) served as a reference standard. Sensitivity, specificity, and positive and negative predictive values were calculated and compared using the Wilcoxon test. Interobserver agreement was determined by calculation of kappa values. Radiation exposure and objective image parameters were assessed and compared with a 2-sided t test. Results: In the reference scan, a total of 94 emboli were detected: 17 in the main and lobar pulmonary level (category A), 47 in the segmental level (category B), and 30 in the subsegmental level (category C). All protocols reached high diagnostic accuracy in the detection of PE in category A. No significant difference was observed between protocols A and B in the detection rate of segmental and subsegmental PE (sensitivity: 93.6% and 91.5% vs 85.1% and 87.2%; positive predictive value: 100% and 97.7% vs 97.5% and 95.3%). Interobserver agreement was excellent at the segmental (kappa = 0.97 and 0.94) and subsegmental levels (kappa = 0.94 and 0.92). Using protocol C, the detection rate of segmental and subsegmental emboli was significantly impaired and interobserver agreement was significantly inferior (sensitivity: 46.8% and 44.6% at segmental and 56.7% and 50.0% at subsegmental level; kappa = 0.70 and 0.60, respectively; P < 0.05). Vessel attenuation was significantly higher in protocol C, [710.3 Hounsfield units (HU) +/- 200.6 SD] than in protocols A (414.4HU +/- 82.5 SD) and B (428.8HU +/- 78.9 SD) (P < 0.001). Signal-to-noise and contrast-to-noise ratios were significantly decreased in protocols B (55.6; 45.8) and C (44.3; 39.4) compared with protocol A (62.1; 52.3) (P < 0.05). Radiation dose was significantly reduced between protocols A and B and protocols B and C [volume CT dose index (CTDIvol): 7.1 vs 4.7 vs 1.4 mGy; dose-length product: 199.5 vs 132.1 vs 39.4mGy.cm and E: 3.39 vs 2.25 vs 0.6 mSv; P < 0.05]. Conclusion: This animal study proves the diagnostic use of a low-dose CTPA protocol with 80 mAs(eff). at 120 kV for PE evaluation without significant loss of diagnostic accuracy compared with a standard protocol with 120mAs(eff). Using an "ultra-low"-dose CTPA protocol (80 kV; 80 mAs(eff).), emboli detection at the segmental and subsegmental level is significantly impaired, whereas evaluation of central emboli is still feasible with high diagnostic accuracy.
引用
收藏
页码:113 / 120
页数:8
相关论文
共 35 条
[1]
FEASIBILITY OF HIGH-RESOLUTION, LOW-DOSE CHEST CT IN EVALUATING THE PEDIATRIC CHEST [J].
AMBROSINO, MM ;
GENIESER, NB ;
ROCHE, KJ ;
KAUL, A ;
LAWRENCE, RM .
PEDIATRIC RADIOLOGY, 1994, 24 (01) :6-10
[2]
Strategies for formulating appropriate MDCT techniques when imaging the chest, abdomen, and pelvis in pediatric patients [J].
Cody, DD ;
Moxley, DM ;
Krugh, KT ;
O'Daniel, JC ;
Wagner, LK ;
Eftekhari, F .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 182 (04) :849-859
[3]
CT angiography in the evaluation of acute pulmonary embolus [J].
Costantino, Mary M. ;
Randall, Geneva ;
Gosselin, Marc ;
Brandt, Marissa ;
Spinning, Kristopher ;
Vegas, C. David .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 191 (02) :471-474
[4]
Radiation dose in helical CT for detection of pulmonary embolism [J].
Diederich, S .
EUROPEAN RADIOLOGY, 2003, 13 (07) :1491-1493
[5]
Non-traumatic thoracic emergencies: CT diagnosis of acute pulmonary embolism: the first 10 years [J].
Ghaye, B ;
Remy, J ;
Remy-Jardin, M .
EUROPEAN RADIOLOGY, 2002, 12 (08) :1886-1905
[6]
Image quality and radiation exposure at pulmonary CT angiography with 100-or 120-kVp protocol: prospective randomized study [J].
Heyer, Christoph M. ;
Mohr, Patrick S. ;
Lemburg, Stefan P. ;
Peters, Soeren A. ;
Nicolas, Volkmar .
RADIOLOGY, 2007, 245 (02) :577-583
[7]
Radiation dose to the female breast from 16-MDCT body protocols [J].
Hurwitz, LM ;
Yoshizumi, T ;
Reiman, RE ;
Paulson, EK ;
Frush, DP ;
Nguyen, GT ;
Toncheva, GI ;
Goodman, PC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) :1718-1722
[8]
Radiation dose from contemporary cardiothoracic multidetector CT Protocols with an anthropomorphic female phantom: Implications for cancer induction [J].
Hurwitz, Lynne M. ;
Reiman, Robert E. ;
Yoshizumi, Terry T. ;
Goodman, Philip C. ;
Toncheva, Greta ;
Nguyen, Giao ;
Lowry, Carolyn .
RADIOLOGY, 2007, 245 (03) :742-750
[9]
The indeterminate CT pulmonary angiogram: Imaging characteristics and patient clinical outcome [J].
Jones, SE ;
Wittram, C .
RADIOLOGY, 2005, 237 (01) :329-337
[10]
Reduced Z-axis coverage multidetector CT angiography for suspected acute pulmonary embolism could decrease dose and maintain diagnostic accuracy [J].
Kallen J.A. ;
Coughlin B.F. ;
O'Loughlin M.T. ;
Stein B. .
Emergency Radiology, 2010, 17 (1) :31-35