Percutaneous fibrin sleeve stripping of failing haemodialysis catheters

被引:23
作者
Johnstone, RD
Stewart, CA
Akoh, JA
Fleet, M
Akyol, M
Moss, JG
机构
[1] Gartnavel Gen Hosp, W Glasgow Hosp Univ NHS Trust, Renal Unit, Glasgow G12 0YN, Lanark, Scotland
[2] Gartnavel Gen Hosp, W Glasgow Hosp Univ NHS Trust, Dept Radiol, Glasgow G12 0YN, Lanark, Scotland
关键词
catheter; fibrin; haemodialysis; sheath; stripping;
D O I
10.1093/ndt/14.3.688
中图分类号
R3 [基础医学]; R4 [临床医学];
学科分类号
1001 ; 1002 ; 100602 ;
摘要
Background. One of the most frequent reasons for failure of haemodialysis lines is catheter blockage caused by fibrin sheath formation. We report our experience of percutaneous fibrin sheath stripping in treating this problem. Methods. A consecutive series of failing haemodialysis catheters underwent percutaneous fibrin sheath stripping in an attempt to retrieve and prolong the life of the catheter. Immediate technical success, clinical success, and primary and secondary patency were measured based on clinical follow-up. Results. Sixteen non-functional permanent haemodialysis lines in 15 patients underwent percutaneous fibrin sheath stripping on 21 occasions. Technical success rate was 100%. Catheter Bow rates sufficient for initial dialysis were achieved in 12 (75%) lines. Successful percutaneous fibrin sheath stripping produced a mean catheter patency of 126 days (range 6-299 days). Conclusions, Percutaneous fibrin sheath stripping is a simple, repeatable procedure that can usefully extend the life of a railing dialysis line. However, a randomized trial will be needed to evaluate its role compared with catheter replacement.
引用
收藏
页码:688 / 691
页数:4
相关论文
共 9 条
[1]  
AKOH JA, 1994, AUT M SCOTT REN ASS
[2]   FIBRIN SHEATH ENTRAPMENT OF PERIPHERALLY INSERTED CENTRAL CATHETERS [J].
CARDELLA, JF ;
LUKENS, ML ;
FOX, PS .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 5 (03) :439-442
[3]   Fibrin sleeve stripping for salvage of failing hemodialysis catheters: Technique and initial results [J].
Crain, MR ;
Mewissen, MW ;
Ostrowski, GJ ;
PazFumagalli, R ;
Beres, RA ;
Wertz, RA .
RADIOLOGY, 1996, 198 (01) :41-44
[4]   5 YEARS EXPERIENCE WITH THE QUINTON PERMCATH FOR VASCULAR ACCESS [J].
GIBSON, SP ;
MOSQUERA, D .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1991, 6 (04) :269-274
[5]   THROMBOSED CENTRAL VENOUS CATHETERS - RESTORING FUNCTION WITH 6-HOUR UROKINASE INFUSION AFTER FAILURE OF BOLUS UROKINASE [J].
HAIRE, WD ;
LIEBERMAN, RP .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1992, 16 (02) :129-132
[6]  
HOSHAL VL, 1971, ARCH SURG-CHICAGO, V102, P353
[7]  
KNELSON MH, 1993, JVIR, V4, P36
[8]   Stripping of failing haemodialysis catheters using the Ampltaz gooseneck snare [J].
Rockall, AG ;
Harris, A ;
Wetton, CWN ;
Taube, D ;
Gedroyc, W ;
AlKutoubi, MA .
CLINICAL RADIOLOGY, 1997, 52 (08) :616-620
[9]  
Shaffer D, 1992, ASAIO J, V38, P55, DOI 10.1097/00002480-199201000-00013