Objectives. The purpose of this study was to describe the Orthopantomograph OP 100 Digipan modification and to determine the image-layer and dosimetric characteristics of this system. Research design. Image-layer contours for the Digipan were determined at resolution limits of 4.0, 3.0, and 1.5 line pairs (Ip) mm(-1) using a resolution grid positioned at intervals along the beam projection paths. A hexagonal test device was placed above the grid to determine the magnification factors at the selected resolution limits. For dosimetry, a 30 mm(2) ionization chamber was placed at the beam entry points while imaging the molar, premolar, and anterior teeth using both the Digipan and conventional film receptors. Results. The maximum resolution exceeded 4.0 Ip mm(-1) in the center of the image layer. Using a 1.5 Ip mm(-1) resolution limit, the focal trough width was 14 mm at 0-degree horizontal angulation, 20 mm at 37 degrees, and 36 mm at 83 degrees. At 1.5 Ip mm(-1), the horizontal magnification was -21% facial and +42% lingual to the center of the image layer in the incisor region; -25% facial and +19% lingual to the center of the image layer at a horizontal angulation of 83 degrees. At 4.0 Ip mm(-1), horizontal magnifications in the incisor region were -2.0% facially and +0.5% lingually; at 83 degrees they were -3.9% facially and +1.1% lingually. The neck entrance dose for imaging the anterior and premolar regions averaged 89.5 (+/-5.6) mu Gy. The cheek entrance doses for imaging the molar region averaged 220.6 (+/-8.0) mu Gy with the Digipan. With film the average entrance doses were 297.9 (+/-8.6) mu Gy (anterior and premolar regions) and 682.1 (+/-16.9) mu Gy (molar region). Conclusions. The Digipan image layer is similar, but slightly narrower than that previously reported for the Orthopantomograph OP 100 using standard film-screen combinations. The resolution at the center of the image layer exceeded 4.0 Ip mm(-1). The Digipan produced images with an entry dose savings of approximately 70% when compared with photo-timed exposures using conventional film with the Orthopantomograph OP 100.
引用
收藏
页码:281 / 287
页数:7
相关论文
共 17 条
[1]  
ARAI Y, 1992, DENTOMAXILLOFAC RAD, V21, P223
[2]   CONSIDERATIONS IN MOVING ELECTRONIC RADIOGRAPHY INTO ROUTINE USE [J].
COX, JR ;
MUKA, E ;
BLAINE, GJ ;
MOORE, SM ;
JOST, RG .
IEEE JOURNAL ON SELECTED AREAS IN COMMUNICATIONS, 1992, 10 (07) :1108-1120
[3]   PRELIMINARY EVALUATION OF A DIGITAL SYSTEM FOR ROTATIONAL PANORAMIC RADIOGRAPHY [J].
DOVE, SB ;
MCDAVID, WD ;
WELANDER, U ;
TRONJE, G .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1992, 73 (05) :623-632
[4]  
ERASO FE, 1996, THESIS U LOUISVILLE
[5]  
HASSFELD S, 1995, CAR, V10, P1002
[6]  
HAYAKAWA Y, 1996, J DENT RES, V75, P367
[7]   DIAGNOSTIC-IMAGING OF DISEASES AFFECTING THE MANDIBLE WITH THE USE OF COMPUTED PANORAMIC RADIOGRAPHY [J].
KASHIMA, I ;
TAJIMA, K ;
NISHIMURA, K ;
YAMANE, R ;
SARAYA, M ;
SASAKURA, Y ;
TAKANO, M .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1990, 70 (01) :110-116
[8]   BONE TRABECULAR PATTERN-ANALYSIS IN DOWN-SYNDROME WITH THE USE OF COMPUTED PANORAMIC RADIOGRAPHY [J].
KASHIMA, I ;
BANDO, S ;
KANISHI, D ;
MIYAKE, K ;
YAMANE, R ;
TAKANO, M .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 1990, 70 (03) :360-364
[9]   COMPUTED RADIOGRAPHY WITH PHOTOSTIMULABLE PHOSPHOR IN ORAL AND MAXILLOFACIAL RADIOLOGY [J].
KASHIMA, I .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1995, 80 (05) :577-598
[10]  
KUNDEL HL, 1986, RADIOL CLIN N AM, V24, P69