Active matrix liquid crystal displays for clinical imaging: Comparison with cathode ray tube displays

被引:25
作者
Pavlicek, W
Owen, JM
Peter, MB
机构
[1] Mayo Clin Scottsdale, Dept Diagnost Radiol, Scottsdale, AZ 85259 USA
[2] Mayo Clin Scottsdale, Dept Clin Engn, Scottsdale, AZ 85259 USA
关键词
D O I
10.1007/BF03167650
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Fifteen large-area, flat-panel displays used for clinical image review were evaluated for image quality and compared with 30 comparably sized cathode ray tube (CRT) monitors. Measurements were of image display patterns by Video Electronic Standards Association (VESA) and a commercial product. Field measurements were made of: maximum and minimum luminance, ambient lighting, characteristic curve (gamma), point shape and size, high-contrast resolution, uniformity, and distortion. Assessments were made of pixel defects, latent image patterns, ghosting artifacts, and viewing angle luminance. Also, a questionnaire was generated for users of the flat-panel and CRT units. Seventeen respondents indicated no preference for either flat panel or CRT. Results show these flat panels to have higher luminance (mean, 177.7 cd/m(2)); larger number of just noticeable differences (JNDs; n = 555), higher gamma, comparable uniformity, and warm-up time. CRTs had less angle viewing dependence and far fewer artifacts (ghosting and latent images). Our questionnaire showed active matrix liquid crystal displays (AMLCD) to be fully acceptable for clinical image viewing. Furthermore, the statistical results show that further testing for new AMLCDs of this type is unwarranted. Copyright (C) 2000 by W.B. Saunders Company.
引用
收藏
页码:155 / 161
页数:7
相关论文
共 25 条
[1]  
*ACR, 1999, AM COLL RAD STAND TE
[2]  
BLUME H, 1997, P RADIOL SOC N AM S, V213, P769
[3]   Theoretical and experimental study of nematic liquid crystal display cells using the in-plane-switching mode [J].
Di Pasquale, F ;
Deng, HF ;
Fernández, FA ;
Day, SE ;
Davies, JB ;
Johnson, MT ;
van der Put, AA ;
van de Eerenbeemd, JMA ;
van Haaren, JAMM ;
Chapman, JA .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1999, 46 (04) :661-668
[4]   Image display for clinicians on medical record workstations [J].
Erickson, BJ ;
Ryan, WJ ;
Gehring, DG ;
Beebe, C .
JOURNAL OF DIGITAL IMAGING, 1997, 10 (03) :38-40
[5]   Performance and function of a desktop viewer at Mayo Clinic Scottsdale [J].
Eversman, WG ;
Pavlicek, W ;
Zavalkovskiy, B ;
Erickson, BJ .
JOURNAL OF DIGITAL IMAGING, 2000, 13 (02) :147-152
[6]   Image quality degradation by light scattering in display devices [J].
Flynn, MJ ;
Badano, A .
JOURNAL OF DIGITAL IMAGING, 1999, 12 (02) :50-59
[7]   TEST PATTERN FOR VIDEO DISPLAYS AND HARD-COPY CAMERAS [J].
GRAY, JE ;
LISK, KG ;
HADDICK, DH ;
HARSHBARGER, JH ;
OOSTERHOF, A ;
SCHWENKER, R .
RADIOLOGY, 1985, 154 (02) :519-527
[8]   Optimization of a contrast-detail-based method for electronic image display quality evaluation [J].
Hangiandreou, NJ ;
Fetterly, KA ;
Felmlee, JP .
JOURNAL OF DIGITAL IMAGING, 1999, 12 (02) :60-67
[9]   Effect of display luminance on the feature detection rates of masses in mammograms [J].
Hemminger, BM ;
Dillon, AW ;
Johnston, RE ;
Muller, KE ;
Deluca, MC ;
Coffey, CS ;
Pisano, ED .
MEDICAL PHYSICS, 1999, 26 (11) :2266-2272
[10]  
*ISO, 1999, 9241 ISO 3