Technical and clinical evaluation of an electrochemistry glucose meter: experience in a diabetes center

被引:13
作者
Chen, HS
Kuo, BI
Hwu, CM
Shih, KC
Kwok, CF
Ho, LT [1 ]
机构
[1] Vet Gen Hosp, Dept Med, Sect Endocrinol & Metab, Taipei, Taiwan
[2] Vet Gen Hosp, Dept Med Res & Educ, Taipei, Taiwan
[3] Vet Gen Hosp, Dept Med, Sect Gen Med, Taipei, Taiwan
[4] Natl Yang Ming Univ, Sch Med, Dept Med, Taipei 112, Taiwan
关键词
self-monitoring of blood glucose; diabetes mellitus; glucose meter; electrochemistry; field test;
D O I
10.1016/S0168-8227(98)00077-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Sensorex (Metertech, Taipei, Taiwan), an electrochemical blood glucose meter, is designed for self-monitoring of blood glucose (BG) concentrations in capillary blood through the use of an electrochemical test strip. The intra-assay coefficients of variation of Sensorex were 5.2, 5.4, and 4.7% at BG levels of 46, 154 and 302 mg/dl respectively. The BG concentrations tested by Sensorex were correlated well with those by YSI method (r congruent to 0.85, P < 0.0001). The intraclass correlation coefficients (r(I)) between the results obtained by Sensorex and YSI were 0.84 in capillary blood and 0.69 in venous whole blood, which indicated good agreement between both methods. The Sensorex was evaluated by error grid analysis and revealed 'acceptance' results. In field test, the Sensorex results obtained by lay users were in concordance with those by trained technicians (r(I) = 0.87). Our data show that the Sensorex glucometer is reliable and easy to use. We also demonstrate a practical clinical approach for the evaluation of a novel SMBG system. (C) 1998 Published by Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 14 条
[1]  
[Anonymous], 1987, DIABETES CARE, V10, P95
[2]   PENTAMIDINE-INDUCED DERANGEMENTS OF GLUCOSE-HOMEOSTASIS - DETERMINANT ROLES OF RENAL-FAILURE AND DRUG ACCUMULATION - A STUDY OF 128 PATIENTS [J].
ASSAN, R ;
MAYAUD, C ;
PERRONNE, C ;
MATHERON, S ;
ASSAN, D ;
ZUCMAN, D ;
CHOTARD, L .
DIABETES CARE, 1995, 18 (01) :47-55
[3]   LABORATORY ASSESSMENT OF GLUCOSE METERS DOES NOT PREDICT RELIABILITY OF CLINICAL-PERFORMANCE [J].
BERNBAUM, M ;
ALBERT, SG ;
MCGINNIS, J ;
BRUSCA, S ;
MILLER, D ;
HOFFMANN, J ;
KAISER, FE ;
MOORADIAN, AD .
LABORATORY MEDICINE, 1994, 25 (01) :32-34
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[6]   WHAT DO STATISTICS REALLY TELL US ABOUT THE QUALITY OF THE DATA FROM SELF-MONITORING OF BLOOD-GLUCOSE [J].
DEDRICK, RF ;
DAVIS, WK .
DIABETIC MEDICINE, 1989, 6 (03) :267-273
[7]  
GRAHAM D, 1985, BIOSENSORS, V1, P161
[8]   SELF-MONITORING OF BLOOD-GLUCOSE BY ADULTS WITH DIABETES IN THE UNITED-STATES POPULATION [J].
HARRIS, MI ;
COWIE, CC ;
HOWIE, LJ .
DIABETES CARE, 1993, 16 (08) :1116-1123
[9]  
KELLY RI, 1990, HUMAN FACTORS SELF M
[10]   NEW APPROACH TO TECHNICAL AND CLINICAL-EVALUATION OF DEVICES FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
KOSCHINSKY, T ;
DANNEHL, K ;
GRIES, FA .
DIABETES CARE, 1988, 11 (08) :619-629