Improved immobilization of 8-hydroxyquinoline on polyacrylonitrile fiber and application of the material to the determination of trace metals in seawater by inductively coupled plasma mass spectrometry

被引:25
作者
Wen, B [1 ]
Shan, X [1 ]
机构
[1] Chinese Acad Sci, Ecoenvironm Sci Res Ctr, Beijing 100085, Peoples R China
关键词
8-hydroxyquinoline immobilized polyacrylonitrile fiber; trace metals; sea water; inductively coupled plasma mass spectrometry;
D O I
10.1007/s00216-002-1551-0
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A modified synthetic method has been developed for immobilization of 8-hydroxyquinoline on polyacrylonitrile fiber. The synthetic conditions, e.g. reagent concentration, reaction temperature and time, were optimized. The features of the newly-modified fiber are higher exchange capacity compared with most other materials containing immobilized 8-hydroxyquinoline, better mechanic characteristics, high stability at both high and low pH, and ease of synthesis. This modified fiber can concentrate and separate trace metals from matrices with higher concentrations of alkali and alkaline earth elements. To validate the characteristics of the modified fiber, trace metals Ag, Be, Cd, Co, Cu, Mn, Ni, Pb, and Zn in the certified reference materials river water (SLRS-4) and seawater (CASS-4, NASS-5, SLEW-3) were pre-concentrated before determination by inductively coupled plasma mass spectrometry. Good agreement was obtained between the data obtained by this method and the certified values.
引用
收藏
页码:948 / 954
页数:7
相关论文
共 29 条
[1]   Determination of cadmium, copper, and zinc by flame atomic absorption spectrometry after preconcentration using a DETA (diethylenetriamine) polymer [J].
Bakircioglu, Y ;
Seren, G ;
Akman, S .
ANALYTICAL LETTERS, 2001, 34 (03) :439-447
[2]   DETERMINATION OF TRACE-METALS IN REFERENCE WATER STANDARDS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY WITH ONLINE PRECONCENTRATION [J].
BEAUCHEMIN, D ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1989, 61 (17) :1857-1862
[3]   SEPARATION OF CR(III) AND CR(VI) USING COMPLEXATION OF CR(III) WITH 8-HYDROXYQUINOLINE AND DETERMINATION OF BOTH SPECIES IN WATERS BY ETA-AAS [J].
BECEIROGONZALEZ, E ;
BARCIELAGARCIA, J ;
BERMEJOBARRERA, P ;
BERMEJOBARRERA, A .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 344 (7-8) :301-305
[4]   DETERMINATION OF TRACE-ELEMENTS IN SEA-WATER BY FLOW-INJECTION ANODIC-STRIPPING VOLTAMMETRY PRECEDED BY IMMOBILIZED QUINOLIN-8-OL SILICA-GEL PRECONCENTRATION [J].
DAIH, BJ ;
HUANG, HJ .
ANALYTICA CHIMICA ACTA, 1992, 258 (02) :245-252
[5]   PRECONCENTRATION AND PURIFICATION OF RARE-EARTH ELEMENTS IN NATURAL-WATERS USING SILICA-IMMOBILIZED 8-HYDROXYQUINOLINE AND A SUPPORTED ORGANOPHOSPHORUS EXTRACTANT [J].
ESSER, BK ;
VOLPE, A ;
KENNEALLY, JM ;
SMITH, DK .
ANALYTICAL CHEMISTRY, 1994, 66 (10) :1736-1742
[6]   SYNTHETIC ASPECTS OF THE CHARACTERIZATION OF SOME SILICA-BOUND COMPLEXING AGENTS [J].
FULCHER, C ;
CROWELL, MA ;
BAYLISS, R ;
HOLLAND, KB ;
JEZOREK, JR .
ANALYTICA CHIMICA ACTA, 1981, 129 (AUG) :29-47
[7]   High-precision measurement of calcium isotopes in carbonates and related materials by multiple collector inductively coupled plasma mass spectrometry (MC-ICP-MS) [J].
Halicz, L ;
Galy, A ;
Belshaw, NS ;
O'Nions, RK .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (12) :1835-1838
[8]   On-line method for inductively coupled plasma mass spectrometric determination of rare earth elements in highly saline brines [J].
Halicz, L ;
Gavrieli, I ;
Dorfman, E .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (09) :811-814
[9]   SILICA-GEL AS AN INSOLUBLE CARRIER FOR PREPARATION OF SELECTIVE CHROMATOGRAPHIC ADSORBENTS - PREPARATION OF 8-HYDROXYQUINOLINE SUBSTITUTED SILICA-GEL FOR CHELATION CHROMATOGRAPHY OF SOME TRACE METALS [J].
HILL, JM .
JOURNAL OF CHROMATOGRAPHY, 1973, 76 (02) :455-458
[10]  
Jarvis K. E., 1992, HDB INDUCTIVELY COUP, DOI [10.1007/BF02290251, DOI 10.1007/BF02290251]