Megavoltage CT image reconstruction during tomotherapy treatments

被引:91
作者
Ruchala, KJ
Olivera, GH
Kapatoes, JM
Schloesser, EA
Reckwerdt, PJ
Mackie, TR
机构
[1] TomoTherapy Inc, Middleton, WI 53562 USA
[2] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
[3] Univ Wisconsin, Dept Human Oncol, Madison, WI 53706 USA
关键词
D O I
10.1088/0031-9155/45/12/303
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An integrated tomotherapy system allows for improved radiotherapy verification by enabling the collection of megavoltage computed tomography (MVCT) images before or after treatment delivery. In this investigation. the possibility of collecting MV tomographic data and reconstructing images during a tomotherapy treatment is examined. By overcoming difficulties with the normalization of modulated treatment data and with the incompleteness of treatment data, it is possible to use data collected during tomotherapeutic treatments for MVCT reconstruction. The benefits of these techniques include potential increases in patient throughput, reductions in imaging dose, visualization of the patient in the treatment position and improvements in image contrast.
引用
收藏
页码:3545 / 3562
页数:18
相关论文
共 38 条
[1]  
Aoki Y, 1990, Radiat Med, V8, P107
[2]  
BALOG J, 1998, THESIS U WISCONSIN M
[3]   First clinical tests using a liquid-filled electronic portal imaging device and a convolution model for the verification of the midplane dose [J].
Boellaard, R ;
van Herk, M ;
Uiterwaal, H ;
Mijnheer, B .
RADIOTHERAPY AND ONCOLOGY, 1998, 47 (03) :303-312
[4]   RADIOTHERAPEUTIC COMPUTED-TOMOGRAPHY WITH SCANNED PHOTON BEAMS [J].
BRAHME, A ;
LIND, B ;
NAFSTADIUS, P .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1987, 13 (01) :95-101
[5]  
GROH BA, 2000, MV KV CONE BEAM CT M, P561
[6]   Feasibility of megavoltage portal CT using an electronic portal imaging device (EPID) and a multi-level scheme algebraic reconstruction technique (MLS-ART) [J].
Guan, HQ ;
Zhu, YP .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :2925-2937
[7]   Tomotherapeutic portal imaging for radiation treatment verification [J].
Hesse, BM ;
Spies, L ;
Groh, BA .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (12) :3607-3616
[8]   Managing geometric uncertainty in conformal intensity-modulated radiation therapy [J].
Jaffray, DA ;
Yan, D ;
Wong, JW .
SEMINARS IN RADIATION ONCOLOGY, 1999, 9 (01) :4-19
[9]   A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets [J].
Jaffray, DA ;
Drake, DG ;
Moreau, M ;
Martinez, AA ;
Wong, JW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (03) :773-789
[10]  
Kapatoes JM, 2000, USE OF COMPUTERS IN RADIATION THERAPY, P294