Dual-pulse lithotripter accelerates stone fragmentation and reduces cell lysis in vitro

被引:27
作者
Sokolov, DL [1 ]
Bailey, MR [1 ]
Crum, LA [1 ]
机构
[1] Univ Washington, Appl Phys Lab, Ctr Ind & Med Ultrasound, Seattle, WA 98105 USA
关键词
lithotripsy; cavitation; hemolysis; shock waves;
D O I
10.1016/S0301-5629(03)00887-1
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Lithotripsy is a common effective treatment for kidney stones. However, focal volumes are often larger than stones, and surrounding tissue is often injured. Our goal was to test in vitro a new lithotripter consisting of two opposing, confocal and simultaneously triggered electrohydraulic sources. The pulses superimpose at the common focus, resulting in pressure doubling and enhanced cavitation growth in a localized, similar to1-cm wide volume. Model gypsum stones and human erythrocytes were exposed to dual pulses or single pulses. At the focus, model stones treated with 100 dual pulses at a charging voltage of 15 kV broke into 8 times the number of fragments as stones treated with 200 single pulses at 18 kV. At axial positions 2 and 4 cm away from the focus, lysis of erythrocytes was reduced or equivalent for dual pulses vs. single pulses. Hence, in half the time, dual pulses increased comminution at the focus without increasing injury in surrounding regions. (E-mail: bailey@apl.washington.edu) (C) 2003 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:1045 / 1052
页数:8
相关论文
共 52 条
[1]   Comparison of electrohydraulic lithotripters with rigid and pressure-release ellipsoidal reflectors. II. Cavitation fields [J].
Bailey, MR ;
Blackstock, DT ;
Cleveland, RO ;
Crum, LA .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 106 (02) :1149-1160
[2]  
BAILEY MR, 1997, ARLTR971 U TEX AUST
[3]  
BAILEY MR, 2001, 17 INT C AC ROM IT
[4]   COLLAPSE OF CAVITATION BUBBLES AND PRESSURES THEREBY PRODUCED AGAINST SOLID BOUNDARIES [J].
BENJAMIN, TB ;
ELLIS, AT .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1966, 260 (1110) :221-&
[5]  
CHAUSSY C, 1980, LANCET, V2, P1265
[6]   A THEORETICAL-STUDY OF CAVITATION GENERATED BY AN EXTRACORPOREAL SHOCK-WAVE LITHOTRIPTER [J].
CHURCH, CC .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1989, 86 (01) :215-227
[7]  
CHURCH CC, 1990, P 12 INT S NONL AC, P433
[8]   Design and characterization of a research electrohydraulic lithotripter patterned after the Dornier HM3 [J].
Cleveland, RO ;
Bailey, MR ;
Fineberg, N ;
Hartenbaum, B ;
Lokhandwalla, M ;
McAteer, JA ;
Sturtevant, B .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (06) :2514-2525
[9]   ACOUSTIC CAVITATION GENERATED BY AN EXTRACORPOREAL SHOCKWAVE LITHOTRIPTER [J].
COLEMAN, AJ ;
SAUNDERS, JE ;
CRUM, LA ;
DYSON, M .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1987, 13 (02) :69-76
[10]   PRESSURE WAVE-FORMS GENERATED BY A DORNIER EXTRACORPOREAL SHOCK-WAVE LITHOTRIPTER [J].
COLEMAN, AJ ;
SAUNDERS, JE ;
PRESTON, RC ;
BACON, DR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1987, 13 (10) :651-657