Design and characterization of a research electrohydraulic lithotripter patterned after the Dornier HM3

被引:80
作者
Cleveland, RO [1 ]
Bailey, MR
Fineberg, N
Hartenbaum, B
Lokhandwalla, M
McAteer, JA
Sturtevant, B
机构
[1] Boston Univ, Dept Aerosp & Mech Engn, Boston, MA 02215 USA
[2] Univ Washington, Appl Phys Lab, Seattle, WA 98105 USA
[3] Indiana Univ, Sch Med, Div Biostat, Indianapolis, IN 46202 USA
[4] H Tech Labs, Santa Monica, CA 90406 USA
[5] CALTECH, Grad Aeronaut Lab, Pasadena, CA 91125 USA
[6] Indiana Univ, Sch Med, Dept Anat & Cell Biol, Indianapolis, IN 46202 USA
关键词
D O I
10.1063/1.1150643
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An electrohydraulic lithotripter has been designed that mimics the behavior of the Dornier HM3 extracorporeal shock wave lithotripter. The key mechanical and electrical properties of a clinical HM3 were measured and a design implemented to replicate these parameters. Three research lithotripters have been constructed on this design and are being used in a multi-institutional, multidisciplinary research program to determine the physical mechanisms of stone fragmentation and tissue damage in shock wave lithotripsy. The acoustic fields of the three research lithotripters and of two clinical Dornier HM3 lithotripters were measured with a PVDF membrane hydrophone. The peak positive pressure, peak negative pressure, pulse duration, and shock rise time of the focal waveforms were compared. Peak positive pressures varied from 25 MPa at a voltage setting of 12 kV to 40 MPa at 24 kV. The magnitude of the peak negative pressure varied from -7 to -12 MPa over the same voltage range. The spatial variations of the peak positive pressure and peak negative pressure were also compared. The focal region, as defined by the full width half maximum of the peak positive pressure, was 60 mm long in the axial direction and 10 mm wide in the lateral direction. The performance of the research lithotripters was found to be consistent at clinical firing rates (up to 3 Hz). The results indicated that pressure fields in the research lithotripters are equivalent to those generated by a clinical HM3 lithotripter. (C) 2000 American Institute of Physics. [S0034-6748(00)01806-2].
引用
收藏
页码:2514 / 2525
页数:12
相关论文
共 33 条
[1]  
AKERS SR, 1989, I TECH, V23, P50
[2]   Modeling of an electrohydraulic lithotripter with the KZK equation [J].
Averkiou, MA ;
Cleveland, RO .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 106 (01) :102-112
[3]   Comparison of electrohydraulic lithotripters with rigid and pressure-release ellipsoidal reflectors. I. Acoustic fields [J].
Bailey, MR ;
Blackstock, DT ;
Cleveland, RO ;
Crum, LA .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1998, 104 (04) :2517-2524
[4]   Comparison of electrohydraulic lithotripters with rigid and pressure-release ellipsoidal reflectors. II. Cavitation fields [J].
Bailey, MR ;
Blackstock, DT ;
Cleveland, RO ;
Crum, LA .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 106 (02) :1149-1160
[5]  
BAILEY MR, 1997, ARLTR971 U TEX AUST
[6]   THE PERFORMANCE OF DIFFERENT PRESSURE PULSE GENERATORS FOR EXTRACORPOREAL LITHOTRIPSY - A COMPARISON BASED ON COMMERCIAL LITHOTRIPTERS FOR KIDNEY-STONES [J].
BUIZZA, A ;
DELLAQUILA, T ;
GIRIBONA, P ;
SPAGNO, C .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (02) :259-272
[7]  
CAMPBELL DS, 1991, J LITHOTR STONE DIS, V3, P147
[8]  
CHAUSSY C, 1980, LANCET, V2, P1265
[9]   MODELING THE DORNIER-HM3-LITHOTRIPTER [J].
CHRISTOPHER, T .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 96 (05) :3088-3095
[10]   A dual passive cavitation detector for localized detection of lithotripsy-induced cavitation in vitro [J].
Cleveland, RO ;
Sapozhnikov, OA ;
Bailey, MR ;
Crum, LA .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 107 (03) :1745-1758