Biochemical oxygen demand sensor using Serratia marcescens LSY 4

被引:37
作者
Kim, MN [1 ]
Kwon, HS [1 ]
机构
[1] Sangmyung Univ, Dept Biol, Seoul 110743, South Korea
关键词
heavy metal ion tolerance; membrane modification; microbial BOD sensor; poly(sodium styrene sulfonate); Serratia marcescens LSY 4;
D O I
10.1016/S0956-5663(98)00107-9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A microbial biochemical oxygen demand (BOD) sensor consisting of Serratia marcescens LSY 4 and an oxygen electrode was prepared for estimation of the biochemical oxygen demand. The response of the BOD sensor was insensitive to pH in the range of pH 6.0 similar to 8.0, and the baseline drift of the signal was nearly absent even in unbuffered aqueous solution. Because heavy metal ions were precipitated from the phosphate buffer solution, unbuffered solution was used to investigate the effect of the concentration of heavy metal ions on the sensor response. Contrary to previous studies, not only Cu2+ and Ag+ but also Cd2+ and Zn2+ significantly decreased the response of the BOD sensor in unbuffered solution. Graft polymerization of sodium styrene sulfonate on the surface of the porous teflon membrane was carried out to absorb the heavy metal ions permeating through the membrane. Tolerance against Zn2+ was induced for S. marcescens LSY 4 to make the cells less sensitive to the presence of heavy metal ions. The membrane modification and the Zn2+ tolerance induction showed some positive effects in such a way that they reduced the inhibitory effects of Zn2+ and Cd2+ on the sensitivity of the BOD sensor. However, they had no effect on the protection of the cells against the interference of Cu2+ and Ag+ on the performance of the sensor. (C) 1999 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 18 条
[1]   Estimation of biochemical oxygen demand in slurry and effluents using ultra-violet spectrophotometry [J].
Brookman, SKE .
WATER RESEARCH, 1997, 31 (02) :372-374
[2]  
GALINDO E, 1992, BIOTECHNOL TECH, V6, P399
[3]   USE OF MICROBIAL ELECTRODES FOR OBSERVATION OF MICROBIOLOGICAL NITROGEN ELIMINATION PROCESS [J].
HIKUMA, M ;
MATSUOKA, H ;
KAWARAI, M ;
KARUBE, I .
BIOTECHNOLOGY AND BIOENGINEERING, 1992, 40 (01) :130-136
[4]   A NOVEL BOD SENSOR BASED ON BACTERIAL LUMINESCENCE [J].
HYUN, CK ;
TAMIYA, E ;
TAKEUCHI, T ;
KARUBE, I .
BIOTECHNOLOGY AND BIOENGINEERING, 1993, 41 (11) :1107-1111
[5]  
*JIS, 1974, K0102 JIS, P33
[6]   BIOCHEMICAL FUEL-CELL UTILIZING IMMOBILIZED CELLS OF CLOSTRIDIUM-BUTYRICUM [J].
KARUBE, I ;
MATSUNAGA, T ;
TSURU, S ;
SUZUKI, S .
BIOTECHNOLOGY AND BIOENGINEERING, 1977, 19 (11) :1727-1733
[7]   MICROBIAL ELECTRODE BOD SENSORS [J].
KARUBE, I ;
MATSUNAGA, T ;
MITSUDA, S ;
SUZUKI, S .
BIOTECHNOLOGY AND BIOENGINEERING, 1977, 19 (10) :1535-1547
[8]   EFFECT OF HEAVY-METAL IONS ON THE EFFICACY OF A MIXED BACILLI BOD SENSOR [J].
LI, F ;
TAN, TC .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (4-5) :315-324
[9]   MONITORING BOD IN THE PRESENCE OF HEAVY-METAL IONS USING A POLY(4-VINYLPYRIDINE)-COATED MICROBIAL SENSOR [J].
LI, F ;
TAN, TC .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (06) :445-455
[10]   STUDY OF BOD MICROBIAL SENSORS FOR WASTE-WATER TREATMENT CONTROL [J].
LI, YR ;
CHU, J .
APPLIED BIOCHEMISTRY AND BIOTECHNOLOGY, 1991, 28-9 :855-863