摘要
The GEANT4 software was developed by RD44, a world-wide collaboration of national institutes, laboratories and large High-Energy Physics experiments. GEANT4 is a public software package composed of tools which can be used to accurately simulate the passage of particles through matter. In this article, the first attempt to use GEANT4 to model a reverse electrode germanium detector (REGe), and to improve also its efficiency calibration procedure, is presented. A variance reduction algorithm based on a directional bias scheme is implemented into GEANT4 in order to accelerate the efficiency computations. A fast optimisation method to model the detector geometry using standard point sources is also presented and validated for point, Marinelli and air filter sources. The simulated full-energy peak efficiencies agreed with the measured values to within 1% between 36 and 1460 keV for these three counting geometries. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:764 / 774
页数:11
相关论文
共 23 条
[1]   VALIDATION OF A METHOD FOR COMPUTER CALCULATION OF GERMANIUM DETECTOR EFFICIENCIES [J].
AALTONEN, H ;
KLEMOLA, S ;
UGLETVEIT, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 339 (1-2) :87-91
[2]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[3]  
[Anonymous], CERNOPEN99121
[4]  
BRAIZINHA B, 2000, P 8 INT C CAL HIGH E
[5]   ROOT - An object oriented data analysis framework [J].
Brun, R ;
Rademakers, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 389 (1-2) :81-86
[6]  
Debertin K., 1988, Gammaand X-Ray Spectrometry with Semiconductor Detectors
[7]  
Firestone R. B., 1997, Table of Isotopes
[8]   Towards a proper modeling of detector and source characteristics in Monte Carlo simulations [J].
García-Talavera, M ;
Neder, H ;
Daza, MJ ;
Quintana, B .
APPLIED RADIATION AND ISOTOPES, 2000, 52 (03) :777-783
[9]   COVARIANCE ANALYSIS AND FITTING OF GERMANIUM GAMMA-RAY DETECTOR EFFICIENCY CALIBRATION DATA [J].
GERALDO, LP ;
SMITH, DL .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 290 (2-3) :499-508
[10]  
GIANI S, CERNOPEN99300