An inductive power link for a wireless endoscope

被引:136
作者
Lenaerts, Bert [1 ]
Puers, Robert [1 ]
机构
[1] Katholieke Univ Leuven, Dept ESAT, MICAS, B-3001 Louvain, Belgium
关键词
wireless endoscopy; capsule endoscopy; transcutaneous link; inductive powering; IVP;
D O I
10.1016/j.bios.2006.06.015
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
An inductive link is presented that can handle freedom of motion. The envisaged application is to power a wireless camera capsule used for non-invasive visual inspection of the small bowel. Up to 150 mW of usable dc power can be delivered to the capsule for the entire duration of its travel along the gastric track. The outer dimensions of the power receiver compartment inside the capsule are theta 10 mm. x 13 mm. The power efficiency of the link is measured to be 1% under worst-case geometrical conditions. The issue of patient's health and safety regarding exposure to the electromagnetic field is addressed. To this purpose, an experimental method is applied to predict specific absorption rates. Conversely, the effect of biological tissue on link performance and practical workability is evaluated as well. Electric shielding of the transmitter coil is applied to reduce biological tissue interaction. In this way, appliance on a living subject is made both safe and practically feasible. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:1390 / 1395
页数:6
相关论文
共 17 条
[1]  
Ahlbom A, 1998, HEALTH PHYS, V74, P494
[2]   A randomized trial comparing wireless capsule endoscopy with push enteroscopy for the detection of small-bowel lesions [J].
Appleyard, M ;
Fireman, Z ;
Glukhovsky, A ;
Jacob, H ;
Shreiver, R ;
Kadirkamanathan, S ;
Lavy, A ;
Lewkowicz, S ;
Scapa, E ;
Shofti, R ;
Swain, P ;
Zaretsky, A .
GASTROENTEROLOGY, 2000, 119 (06) :1431-1438
[3]  
Arena A, 2005, STUD HEALTH TECHNOL, V114, P167
[4]   A prospective trial comparing small bowel radiographs and video capsule endoscopy for suspected small bowel disease [J].
Costamagna, G ;
Shah, SK ;
Riccioni, ME ;
Foschia, F ;
Mutignani, M ;
Perri, V ;
Vecchioli, A ;
Brizi, MG ;
Picciocchi, A ;
Marano, P .
GASTROENTEROLOGY, 2002, 123 (04) :999-1005
[5]   Diagnosing small bowel Crohn's disease with wireless capsule endoscopy [J].
Fireman, Z ;
Mahajna, E ;
Broide, E ;
Shapiro, M ;
Fich, L ;
Sternberg, A ;
Kopelman, Y ;
Scapa, E .
GUT, 2003, 52 (03) :390-392
[6]   The dielectric properties of biological tissues .1. Literature survey [J].
Gabriel, C ;
Gabriel, S ;
Corthout, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2231-2249
[7]   The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2251-2269
[8]   Wireless capsule endoscopy [J].
Iddan, G ;
Meron, G ;
Glukhovsky, A ;
Swain, P .
NATURE, 2000, 405 (6785) :417-417
[9]  
KAZIMIERCZUK MK, 1995, RESONANT POWER CONVE, P347
[10]   DESIGN OF RADIO-FREQUENCY POWERED COILS FOR IMPLANT INSTRUMENTS [J].
KO, WH ;
LIANG, SP ;
FUNG, CDF .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1977, 15 (06) :634-640