Spinal stability with anterior or posterior Ray threaded fusion cages

被引:20
作者
Hitchon, PW
Goel, V
Rogge, T
Dooris, A
Drake, J
Torner, J
机构
[1] Univ Iowa, Coll Med, Dept Biomed Engn, Div Neurosurg, Iowa City, IA USA
[2] Univ Iowa, Coll Med, Sch Prevent Med, Iowa City, IA USA
[3] Vet Adm Med Ctr, Iowa City, IA USA
关键词
spinal fusion; spinal stabilization; threaded fusion cage; biomechanical study;
D O I
10.3171/spi.2000.93.1.0102
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. The authors conducted a study to determine if the rigidity supplied to the spine by posterior placement of the Ray threaded fusion cage (TFC) is further enhanced by the placement of pedicle screws and, additionally, if bilateral anteriorly placed TFCs render the spine more rigid than a single anteriorly placed TFC. Methods. Ten human cadaveric spinal specimens (L2-S1) were affixed within a testing frame. Loads of 1.5, 3, 4.5, and 6 Nm were applied to the spine in six degrees of freedom: flexion-extension, right and left lateral bending, and right and left axial rotation. Motion in an x, y, and z cartesian axis system was tracked using dual video cameras following light-emitting diodes attached to the spine and base plate. Load testing of the spines was performed in the intact mode, following which the spinal segments were randomized to receive anterior or posterior instrumentation. In five spine specimens we performed posterior discectomy, posterior lumbar interbody fusion (PLIF) with placment of femoral rings and pedicle screws, PLIF with bilateral TFCs, and bilateral TFCs with pedicle screws. Five other spines underwent anterior-approach discectomy, followed by implantation of a unilateral cage and bilateral cages. Load testing was performed after each step. Conclusion. Spines in which PLIF with pedicle screws and TFCs with pedicle screws were placed were more rigid than after discectomy in all directions of motion except flexion. Anterior discectomy provided significantly (p less than or equal to 0.05) less stability in left and right axial rotation than the intact spines and following posterior discectomy. Following anterior implantation of bilateral TFCs, spines were significantly more rigid than after discectomy in all directions except extension.
引用
收藏
页码:102 / 108
页数:7
相关论文
共 21 条
[1]
Brantigan JW, 1998, TECH NEUROSURG, V4, P216
[2]
A CARBON-FIBER IMPLANT TO AID INTERBODY LUMBAR FUSION - 2-YEAR CLINICAL-RESULTS IN THE 1ST 26 PATIENTS [J].
BRANTIGAN, JW ;
STEFFEE, AD .
SPINE, 1993, 18 (14) :2106-2117
[3]
Posterior lumbar interbody fusion - A biomechanical comparison, including a new threaded cage [J].
Brodke, DS ;
Dick, JC ;
Kunz, DN ;
McCabe, R ;
Zdeblick, TA .
SPINE, 1997, 22 (01) :26-31
[4]
BIOMECHANICS OF SEQUENTIAL POSTERIOR LUMBAR SURGICAL ALTERATIONS [J].
CUSICK, JF ;
YOGANANDAN, N ;
PINTAR, FA ;
REINARTZ, JM .
JOURNAL OF NEUROSURGERY, 1992, 76 (05) :805-811
[6]
DICKMAN CA, 1994, SPINE, V19, pS2266
[7]
GARFIN SR, 1994, SPINE, V19, pS2300
[8]
KINEMATICS OF THE WHOLE LUMBAR SPINE - EFFECT OF DISCECTOMY [J].
GOEL, VK ;
GOYAL, S ;
CLARK, C ;
NISHIYAMA, K ;
NYE, T .
SPINE, 1985, 10 (06) :543-554
[9]
GOEL VK, 1990, BIOMECHANICS SPINE C, P97
[10]
Biomechanical studies on two anterior thoracolumbar implants in cadaveric spines [J].
Hitchon, PW ;
Goel, VK ;
Rogge, T ;
Grosland, NM ;
Torner, J .
SPINE, 1999, 24 (03) :213-218