Effect of long-term physiological activity on the long-term stem stability of cemented hip arthroplasty: in vitro comparison of three commercial bone cements

被引:3
作者
Bialoblocka-Juszczyk, E. [1 ,2 ]
Cristofolini, L. [1 ,2 ]
Erani, P. [1 ]
Viceconti, M. [1 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, Fac Ingn, Bologna, Italy
关键词
cemented hip prostheses; acrylic bone cement; long-term stability; fatigue testing; physiological loading; pre-clinical validation; aseptic loosening of the stem; PRECLINICAL ASSESSMENT; FAILURE; VALIDATION; MIGRATION; BONELOC(R); COMPONENTS; ENDURANCE; MANTLE; DAMAGE; FEMUR;
D O I
10.1243/09544119JEIM641
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
Long-term endurance of the cement mantle is fundamental for the survival of cemented hip prostheses. Current protocols to characterize bone cements are unsuitable to predict the actual clinical outcome. The aim of this study was to assess if it is possible to rank cement types having diverse clinical outcome by using a simplified in vitro physiological test. Composite femurs were implanted with identical stems (Lubinus-SPII), using different commercial cement types: CMW1 to represent cement with poor clinical outcome; Simplex-P and Cemex-RX to represent cements with a positive clinical outcome. Implanted femurs were subjected to a validated protocol that simulated a demanding but physiological loading spectrum. Inducible micromotions and permanent migrations were recorded throughout the test. After test completion, the cement mantles were sectioned and inspected with dye penetrants to quantify the fatigue-induced cracks. Micromotions did not differ significantly between cement types (possibly because a successful prosthesis was chosen that is very stable in the host bone). Significant differences were observed in terms of cement cracks: CMW1 induced significantly more numerous and larger cracks than Simplex-P and Cemex-RX; no difference was observed between Simplex-P and Cemex-RX. This indicates that this protocol: (a) can discriminate between 'good' and 'bad' cements and (b) yields consistent results when comparable cements are tested. The proposed protocol overcomes the limitations of existing standardized material tests for bone cements. New cements can be assessed in comparison with other cements with known (positive/negative) clinical outcome, tested with the same protocol.
引用
收藏
页码:53 / 65
页数:13
相关论文
共 47 条
[1]
ASEPTIC LOOSENING OF CEMENT-ANCHORED TOTAL HIP REPLACEMENTS [J].
ALDINGER, G ;
GEKELER, J .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1982, 100 (01) :19-25
[2]
[Anonymous], 1994, NIH Consens Statement, V12, P1
[3]
Preclinical testing of femoral hip components: An experimental investigation with four prostheses [J].
Britton, JR ;
Prendergast, PJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (05) :872-880
[4]
Mechanical simulation of muscle loading on the proximal femur: analysis of cemented femoral component migration with and without muscle loading [J].
Britton, JR ;
Walsh, LA ;
Prendergast, PJ .
CLINICAL BIOMECHANICS, 2003, 18 (07) :637-646
[5]
Preclinical assessment of the long-term endurance of cemented hip stems. Part 1: effect of daily activities a comparison of two load histories [J].
Cristofolini, L. ;
Teutonico, A. Saponara ;
Savigni, P. ;
Erani, P. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H6) :569-584
[6]
Preclinical assessment of the long-term endurance of cemented hip stems.: Part 2:: in-vitro and ex-vivo fatigue damage of the cement mantle [J].
Cristofolini, L. ;
Erani, P. ;
Savigni, P. ;
Bordini, B. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H6) :585-599
[7]
A critical analysis of stress shielding evaluation of hip prostheses [J].
Cristofolini, L .
CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 1997, 25 (4-5) :409-483
[8]
Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[9]
Comparative in vitro study on the long term performance of cemented hip stems: validation of a protocol to discriminate between "good" and "bad" designs [J].
Cristofolini, L ;
Teutonico, AS ;
Monti, L ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1603-1615
[10]
Development and validation of a technique for strain measurement inside polymethyl methacrylate [J].
Cristofolini, L ;
Viceconti, M .
JOURNAL OF STRAIN ANALYSIS FOR ENGINEERING DESIGN, 2000, 35 (01) :21-33