Fiske modes in superconducting tunnel junction detectors

被引:12
作者
Friedrich, S
Cunningham, MF
Frank, M
Labov, SE
Barfknecht, AT
Cramer, SP
机构
[1] Univ Calif Lawrence Livermore Natl Lab, Livermore, CA 94550 USA
[2] Univ Calif Berkeley, Lawrence Berkeley Lab, Berkeley, CA 94720 USA
[3] Conductus Inc, Sunnyvale, CA 94086 USA
基金
美国国家航空航天局; 美国国家卫生研究院;
关键词
cryogenic X-ray detectors; superconducting tunnel junctions; Fiske modes;
D O I
10.1016/S0168-9002(99)01348-0
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We are developing superconducting tunnel junction (STJ) X-ray spectrometers with high-energy resolution( < 10 eV for keV X-rays) and high count rate capabilities ( > 10 000 counts/s). We have examined the influence of Fiske mode resonances on energy resolution, maximum count rate and area of Nb-Al-AlOx-Al-Nb STJ detectors. At low count rate, these detectors have an energy resolution of 2.3 eV FWHM at 70 eV. At 20 000 counts/s, the resolution is 8.2 eV FWHM at 277 eV. We observe the optimum detector performance when the influence of Fiske modes is minimized. (C) 2000 Published by Elsevier Science B.V. All rights reserved.
引用
收藏
页码:151 / 155
页数:5
相关论文
共 20 条
[1]   High energy resolution bolometers for nuclear physics and x-ray spectroscopy [J].
Alessandrello, A ;
Beeman, JW ;
Brofferio, C ;
Cremonesi, O ;
Fiorini, E ;
Giuliani, A ;
Haller, EE ;
Monfardini, A ;
Nucciotti, A ;
Pavan, M ;
Pessina, G ;
Previtali, E ;
Zanotti, L .
PHYSICAL REVIEW LETTERS, 1999, 82 (03) :513-515
[2]  
Barfknetcht A. T., 1991, IEEE T MAG, VMAG-27, P970
[3]  
Benner WH, 1997, J AM SOC MASS SPECTR, V8, P1094
[4]   Superconducting particle detectors [J].
Booth, NE ;
Goldie, DJ .
SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 1996, 9 (07) :493-516
[5]   QUASI-PARTICLE TRAPPING AND THE QUASI-PARTICLE MULTIPLIER [J].
BOOTH, NE .
APPLIED PHYSICS LETTERS, 1987, 50 (05) :293-295
[6]  
ECK RE, 1964, P 9 INT C LOW TEMP P
[7]   TEMPERATURE + MAGNETIC FIELD DEPENDENCES OF JOSEPHSON TUNNELING CURRENT [J].
FISKE, MD .
REVIEWS OF MODERN PHYSICS, 1964, 36 (1P1) :221-&
[8]   Cryogenic high-resolution X-ray spectrometers for SR-XRF and microanalysis [J].
Frank, M ;
Mears, CA ;
Labov, SE ;
Hiller, LJ ;
le Grand, JB ;
Lindeman, MA ;
Netel, H ;
Chow, D ;
Barfknecht, AT .
JOURNAL OF SYNCHROTRON RADIATION, 1998, 5 :515-517
[9]   Energy resolution and high count rate performance of superconducting tunnel junction x-ray spectrometers [J].
Frank, M ;
Hiller, LJ ;
le Grand, JB ;
Mears, CA ;
Labov, SE ;
Lindeman, MA ;
Netel, H ;
Chow, D ;
Barfknecht, AT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (01) :25-31
[10]   Experimental quasiparticle dynamics in a superconducting, imaging x-ray spectrometer [J].
Friedrich, S ;
Segall, K ;
Gaidis, MC ;
Wilson, CM ;
Prober, DE ;
Szymkowiak, AE ;
Moseley, SH .
APPLIED PHYSICS LETTERS, 1997, 71 (26) :3901-3903