Experimental results on fast 2D-encoded magnetic particle imaging

被引:136
作者
Gleich, B. [1 ]
Weizenecker, J. [1 ]
Borgert, J. [1 ]
机构
[1] Philips Res Europe Hamburg, Sector Med Imaging Syst, D-22335 Hamburg, Germany
关键词
Optical resolving power - Semiconductor quantum dots - Two dimensional;
D O I
10.1088/0031-9155/53/6/N01
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
This paper presents the first experimental results on magnetic particle imaging with full 2D encoding. The encoding speed achieved was 3.88 ms for a field of view of 1 x 1 cm(2). Small phantoms composed of several dots each filled with 200 nl undiluted Resovist ( 500 mmol( Fe) l(-1)) were scanned. A resolution of better than 1 mm was achieved for a frame rate of 25 frames s(-1).
引用
收藏
页码:N81 / N84
页数:4
相关论文
共 6 条
[1]
Chikazumi S., 1964, PHYS MAGNETISM
[2]
Limits of detection of SPIO at 3.0 T using T2* relaxometry [J].
Dahnke, H ;
Schaeffter, T .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (05) :1202-1206
[3]
Tomographic imaging using the nonlinear response of magnetic particles [J].
Gleich, B ;
Weizenecker, R .
NATURE, 2005, 435 (7046) :1214-1217
[4]
Magnetic iron oxide particles coated with carboxydextran for parenteral administration and liver contrasting - Pre-clinical profile of SH U555A [J].
Lawaczeck, R ;
Bauer, H ;
Frenzel, T ;
Hasegawa, M ;
Ito, Y ;
Kito, K ;
Miwa, N ;
Tsutsui, H ;
Volger, H ;
Weinmann, HJ .
ACTA RADIOLOGICA, 1997, 38 (04) :584-597
[5]
VLAARDINGERBROE.MT, 2003, MAGNETIC RESONANCE I
[6]
A simulation study on the resolution and sensitivity of magnetic particle imaging [J].
Weizenecker, J. ;
Borgert, J. ;
Gleich, B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (21) :6363-6374