Development of a new biomimetic sensor based on an FeIIIFeII complex for the determination of phenolic compounds

被引:15
作者
de Oliveira, Ines Rosane W. Z. [1 ]
Neves, Ademir [2 ]
Vieira, Iolanda Cruz [1 ]
机构
[1] Univ Fed Santa Catarina, Dept Quim, BR-88040900 Florianopolis, SC, Brazil
[2] LABINC, BR-88040900 Florianopolis, SC, Brazil
来源
SENSORS AND ACTUATORS B-CHEMICAL | 2008年 / 129卷 / 01期
关键词
biomimetic sensor; phenolic compound; pharmaceutical formulation;
D O I
10.1016/j.snb.2007.08.047
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The dinuclear iron complex, [(FeFeII)-Fe-III(BPBPMP)(OAc)(2)]ClO4, containing the unsymmetrical ligand, 2-bis[{(2-pyridylmethyl)-aminomethyl}-6-{(2-hydroxybenzyl)(2-pyridylmethyl)}-aminomethyl]-4-methylphenol (H2BPBPMP), has been described as being the first synthetic analogue for the redox properties of uteroferrin. This iron complex was employed in the construction of a new biomimetic sensor for the determination of phenolic compounds by square-wave voltammetry. The responses under optimized conditions were obtained for a 75:15:10% (w/w/w) composition of the graphite powder:Nujol:iron complex, 0.1 mol L-1 phosphate buffer solution (pH 7.0) and 4.5 x 10(-5) mol L-1 hydrogen peroxide. From the several phenolic compounds available (carbidopa, catechol, dopamine, hydroquinone, L-dopa and methyldopa), dopamine was selected, used for optimization of the biomimetic sensor and applied in determinations in pharmaceutical formulations. The analytical curve was linear for dopamine concentrations from 5.0 x 10(-5) to 6.5 x 10(-3) mol L-1 (r = 0.9993). The recovery of dopamine from pharmaceutical sample ranged from 95.8 to 104.2% and the relative standard deviation was 2.6% for a solution containing 3.0 x 10(-3) mol L-1 dopamine (n = 12). The lifetime of this biomimetic sensor was 6 months (at least 800 determinations). The results obtained for dopamine in pharmaceutical samples using the proposed biomimetic sensor and those obtained by the official method are in agreement at the 95% confidence level. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:424 / 430
页数:7
相关论文
共 32 条
[1]   Determination of epinephrine and dopamine in pharmaceutical formulations using a biosensor based on carbon paste modified with crude extract of cara root (Dioscorea bulbifera) [J].
Caruso, CS ;
Vieira, IC ;
Fatibello-Filho, O .
ANALYTICAL LETTERS, 1999, 32 (01) :39-50
[2]   Enzymatic mechanisms of phosphate and sulfate transfer [J].
Cleland, W. Wallace ;
Hengge, Alvan C. .
CHEMICAL REVIEWS, 2006, 106 (08) :3252-3278
[3]   Biomimetic sensor based on a novel copper complex for the determination of hydroquinone in cosmetics [J].
de Oliveira, Ines Rosane W. Z. ;
Osoria, Renata El-Hage M. de Barros ;
Neves, Ademir ;
Vieira, Iolanda Cruz .
SENSORS AND ACTUATORS B-CHEMICAL, 2007, 122 (01) :89-94
[4]   Immobilization procedures for the development of a biosensor for determination of hydroquinone using chitosan and gilo (Solanum gilo) [J].
de Oliveira, IRWZ ;
Vieira, IC .
ENZYME AND MICROBIAL TECHNOLOGY, 2006, 38 (3-4) :449-456
[5]   Biosensor based on chitosan biopolymer and crude extract of ginger (Zingiber officinales rosc.) for the determination of hydroquinone in wastewater of photographic process [J].
de Oliveira, IRWZ ;
Vieira, IC ;
Lupetti, KO ;
Fatibello-Filho, O ;
de Fávere, VT ;
Laranjeira, MCM .
ANALYTICAL LETTERS, 2004, 37 (15) :3111-3127
[6]  
de Oliveira IRWZ.FernandesSC, 2007, J PHARM BIOMED ANAL, V40, P661
[7]   Synthesis, characterization and structure of a new zinc(II) complex containing the hexadentate N,N′,N,N′-bis[(2-hydroxy-3,5-di-tert-butylbenzyl) (2-pyridylmethyl)]-ethylenediamine ligand:: Generation of phenoxyl radical species [J].
dos Anjos, A ;
Bortoluzzi, AJ ;
Szpoganicz, B ;
Caro, MSB ;
Friedermann, GR ;
Mangrich, AS ;
Neves, A .
INORGANICA CHIMICA ACTA, 2005, 358 (11) :3106-3114
[8]   A new dinuclear unsymmetric copper(II) complex as model for the active site of catechol oxidase [J].
Fernandes, C ;
Neves, A ;
Bortoluzzi, AJ ;
Mangrich, AS ;
Rentschler, E ;
Szpoganicz, B ;
Schwingel, E .
INORGANICA CHIMICA ACTA, 2001, 320 (1-2) :12-21
[9]   Langmuir-Blodgett films from polyaniline/ruthenium complexes as modified electrodes for detection of dopamine [J].
Ferreira, M ;
Dinelli, LR ;
Wohnrath, K ;
Batista, AA ;
Oliveira, ON .
THIN SOLID FILMS, 2004, 446 (02) :301-306
[10]   Ion selectivity of a poly(3-pentylmethoxythiophene) LB-layer modified carbon-fiber microelectrode as a consequence of the second order filtering in voltcoulometry [J].
Gmucova, Katarina ;
Weis, Martin ;
Barancok, Drahoslav ;
Cirak, Julius ;
Tomcik, Pavol ;
Pavlasek, Juraj .
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 2007, 70 (03) :385-390