Finite element study of trochanteric gamma nail for trochanteric fracture

被引:151
作者
Sitthiseripratip, K [1 ]
Van Oosterwyck, H
Vander Sloten, J
Mahaisavariya, B
Bohez, ELJ
Suwanprateeb, J
Van Audekercke, R
Oris, P
机构
[1] Asian Inst Technol, Pathum Thani, Thailand
[2] Katholieke Univ Leuven, Div Biomech & Engn Design, B-3001 Louvain, Belgium
[3] Mahidol Univ, Siriraj Hosp, Fac Med, Bangkok 10700, Thailand
[4] Natl Met & Mat Technol Ctr, Pathum Thani, Thailand
关键词
trochanteric fracture; trochanteric gamma nail (TGN); finite element analysis;
D O I
10.1016/S1350-4533(02)00185-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
A three-dimensional finite element study of trochanteric fracture fixation by a trochanteric gamma nail (TGN) was investigated in this study. The analyses were performed under one-legged stance load boundary conditions to study the stress distribution and displacements. The influence of material properties (E-modulus) of the implant, the bone and contact condition in the fracture zone was determined. The results show that the stresses in the implant were lower in case of titanium alloy implant material but at the same time higher displacements occurred. The results also indicate that the stresses in the TGN gradually reduced throughout the healing process of the bone in the fracture zone. (C) 2003 IPEM. Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:99 / 106
页数:8
相关论文
共 13 条
[1]
The effect of sandblasting treatment on endurance properties of titanium alloy hip prostheses [J].
Baleani, M ;
Viceconti, M ;
Toni, A .
ARTIFICIAL ORGANS, 2000, 24 (04) :296-299
[2]
BLACK J, 1998, HDB BIOMATERIALS PRO
[3]
GROSSE A, 1995, AIOD NEWSLETTER
[4]
Failure of femoral head fixation: a cadaveric analysis of lag screw cut-out with the gamma locking nail and AO dynamic hip screw [J].
Haynes, RC ;
Poll, RG ;
Miles, AW ;
Weston, RB .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1997, 28 (5-6) :337-341
[5]
Cutting-out of the lag screw after internal fixation with the Asiatic gamma nail [J].
Kawaguchi, S ;
Sawada, K ;
Nabeta, Y .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1998, 29 (01) :47-53
[6]
Geometric mismatch of the gamma nail to the Chinese femur [J].
Leung, KS ;
Procter, P ;
Robioneck, B ;
Behrens, K .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (323) :42-48
[7]
Multicenter trial of modified gamma nail in East Asia [J].
Leung, KS ;
Chen, CM ;
So, WS ;
Sato, K ;
Lai, CH ;
Machaisavariya, B ;
Suntharalingam, S .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (323) :146-154
[8]
SITTHISERIPRATI.K, 2000, THESIS KU LEUVEN
[9]
SITTHISERIPRATI.K, 2001, P 18 C INT SOC BIOM
[10]
Techanivete P, 1995, J THAI ORTHOP ASS, V20, P2