Osteollysis caused by tibial component debonding in total knee arthroplasty

被引:13
作者
Cheng, Kevin
Pruitt, Lisa
Zaloudek, Charles
Ries, Michael D.
机构
[1] Univ Calif San Francisco, Dept Orthopaed Surg, San Francisco, CA 94143 USA
[2] Univ Calif Berkeley, Dept Bioengn, Berkeley, CA 94720 USA
[3] Univ Calif San Francisco, Dept Pathol, San Francisco, CA 94143 USA
关键词
D O I
10.1097/01.blo.0000196044.42413.c7
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Late failure of total knee arthroplasties usually results from ultrahigh molecular weight polyethylene wear or implant loosening. Early failure from osteolysis is uncommon. However, we treated a patient with a failed total knee arthroplasty from osteolysis that developed 2 years postoperatively. The failure was associated with tibial component debonding from the cement mantle with abundant cement and metal debris. Although there was some third-body debris in the ultrahigh molecular weight polyethylene insert surface, the insert wear was not extensive. Although abundant cement and metal debris were found in the periarticular soft tissues, no ultrahigh molecular weight polyethylene was seen in histologic specimens under polarized light. The osteolysis seems to have been caused primarily by debris generated from debonding and torsional motion at the tibial baseplate-cement interface rather than the bearing surface. Although this failure mechanism has been well recognized in cemented total hip arthroplasties, it has not been reported to be a substantial cause of failure in total knee arthroplasties.
引用
收藏
页码:333 / 336
页数:4
相关论文
共 18 条
[1]
The Insall-Burstein total knee replacement in osteoarthritis - A 10-year minimum follow-up [J].
Aglietti, P ;
Buzzi, R ;
De Felice, R ;
Giron, F .
JOURNAL OF ARTHROPLASTY, 1999, 14 (05) :560-565
[2]
THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[3]
Tibial component failure mechanisms in total knee arthroplasty [J].
Berend, ME ;
Ritter, MA ;
Meding, JB ;
Faris, PM ;
Keating, EM ;
Redelman, R ;
Faris, GW ;
Davis, KE .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) :26-34
[4]
Davey J R, 1993, J Arthroplasty, V8, P23
[5]
Total knee replacement in young, active patients - Long-term follow-up and functional outcome [J].
Diduch, DR ;
Insall, JN ;
Scott, WN ;
Scuderi, GR ;
FontRodriguez, D .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (04) :575-582
[6]
A critical assessment of proximal macrotexturing on cemented femoral components [J].
Duffy, GP ;
Muratoglu, OK ;
Biggs, SA ;
Larson, SL ;
Lozynsky, AJ ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2001, 16 (08) :42-48
[7]
Initial stability of cemented femoral stems as a function of surface finish, collar, and stem size [J].
Ebramzadeh, E ;
Sangiorgio, SN ;
Longjohn, DB ;
Buhari, CE ;
Dorr, LD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (01) :106-115
[8]
The effect of malalignment on stresses in polyethylene component of total knee prostheses - a finite element analysis [J].
Liau, JJ ;
Cheng, CK ;
Huang, CH ;
Lo, WH .
CLINICAL BIOMECHANICS, 2002, 17 (02) :140-146
[9]
The Swedish total hip replacement register [J].
Malchau, H ;
Herberts, P ;
Eisler, T ;
Garellick, G ;
Söderman, P .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A :2-20
[10]
In vitro measurement of strain in the bone cement surrounding the femoral component of total hip replacements during simulated gait and stair-climbing [J].
OConnor, DO ;
Burke, DW ;
Jasty, M ;
Sedlacek, RC ;
Harris, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1996, 14 (05) :769-777