Adjacent Vertebral Body Fracture Following Vertebroplasty With Polymethylmethacrylate or Calcium Phosphate Cement Biomechanical Evaluation of the Cadaveric Spine

被引:88
作者
Nouda, Shinya [1 ]
Tomita, Seiji [1 ]
Kin, Akihiro [1 ]
Kawahara, Kunihiko [1 ]
Kinoshita, Mitsuo [1 ]
机构
[1] Osaka Med Coll, Dept Orthoped Surg, Osaka, Japan
关键词
vertebral body fracture; vertebroplasty; biomechanics; polymethylmethacrylate; calcium phosphate cement; PERCUTANEOUS TRANSPEDICULAR VERTEBROPLASTY; COMPRESSION; KYPHOPLASTY;
D O I
10.1097/BRS.0b013e3181abc150
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
Study Design. A biomechanical study using human cadaveric thoracolumbar spinal columns. Objective. To compare the effect of treatment by vertebroplasty (VP) with polymethylmethacrylate cement and VP with calcium phosphate cement on the creation of adjacent vertebral body fracture following VP. Summary of Background Data. Adjacent vertebral body fractures have been reported as a complication followingVP. Methods. Twenty-four spinal columns (T10-L2) from human cadavers were subjected to dual energy radiograph absorptiometry to assess bone mineral density. They were divided into the P group and C group, and experimental vertebral compression fractures were created at T12 vertebrae. T12 vertebrae were augmented with polymethylmethacrylate and calcium phosphate cement in the P group and C group, respectively. Each spinal column was compressed until a new fracture occurred at any vertebra, and the location of newly fractured vertebra and failure load was investigated. Results. There was no significant difference in bone mineral density at each level within each group. In the P group, a new fracture occurred at T10 in 2 specimens, T11 in 8, and L1 in 2. In the C group, it occurred at T10 in 1 specimen, T11 in 2, L1 in 1, and T12 (treated vertebra) in 8. The failure loads of the spinal column were 1774.8 +/- 672.3 N and 1501.2 +/- 556.5 N in the P group and C group, respectively. There was no significant difference in the failure load of the spinal column between each group. Conclusion. New vertebral fractures occurred at the vertebra adjacent to augmented vertebrae in the P group and in the augmented vertebrae in the C group. The difference in the fractured site may be because of the difference in strength between the 2 bone filler materials. Therefore, the strength of bone filler materials is considered a risk factor in developing adjacent vertebral body fractures after VP.
引用
收藏
页码:2613 / 2618
页数:6
相关论文
共 29 条
[1]
The biomechanics of vertebroplasty - The effect of cement volume on mechanical behavior [J].
Belkoff, SM ;
Mathis, JM ;
Jasper, LE ;
Deramond, H .
SPINE, 2001, 26 (14) :1537-1541
[2]
An ex vivo biomechanical evaluation of a hydroxyapatite cement for use with vertebroplasty [J].
Belkoff, SM ;
Mathis, JM ;
Jasper, LE ;
Deramond, H .
SPINE, 2001, 26 (14) :1542-1546
[3]
CHEUNG KMC, 2005, SPINE, V30, P84
[4]
Temperature elevation caused by bone cement polymerization during vertebroplasty [J].
Deramond, H ;
Wright, NT ;
Belkoff, SM .
BONE, 1999, 25 (02) :17S-21S
[5]
Ebara Sohei, 2004, Clin Calcium, V14, P99
[6]
FORMALIN FIXATION EFFECTS ON VERTEBRAL BONE-DENSITY AND FAILURE MECHANICS - AN IN-VITRO STUDY OF HUMAN AND SHEEP VERTEBRAE [J].
EDMONDSTON, SJ ;
SINGER, KP ;
DAY, RE ;
BREIDAHL, PD ;
PRICE, RI .
CLINICAL BIOMECHANICS, 1994, 9 (03) :175-179
[7]
GALIBERT P, 1987, NEUROCHIRURGIE, V33, P166
[8]
Percutaneous transpedicular vertebroplasty with PMMA:: operative technique and early results -: A prospective study for the treatment of osteoporotic compression fractures [J].
Heini, PF ;
Wälchli, B ;
Berlemann, U .
EUROPEAN SPINE JOURNAL, 2000, 9 (05) :445-450
[9]
Progressive, repeated lumbar compression fracture at the same level after vertebral kyphoplasty with calcium phosphate cement - Case report [J].
Heo, Dong-Hwa ;
Kuh, Sung-Uk .
JOURNAL OF NEUROSURGERY-SPINE, 2007, 6 (06) :559-562
[10]
HIRANO M, 1993, NEW CERAMICS, V5, P55