Cap image - A newly-developed computer-aided videoframe analysis system for dynamic capillaroscopy

被引:233
作者
Klyscz, T
Junger, M
Jung, F
Zeintl, H
机构
[1] INST HERZ & KREISLAUFPHYSIOL,DRESDEN,GERMANY
[2] INGN DR H ZEINTL,HEIDELBERG,GERMANY
来源
BIOMEDIZINISCHE TECHNIK | 1997年 / 42卷 / 06期
关键词
videoframe analysis; digital image processing; capillary microscopy; red blood cell velocity; microcirculation; computer-aided diagnosis; cap image;
D O I
10.1515/bmte.1997.42.6.168
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We describe a newly developed multi-function video image analysis system for the computer-aided evaluation of capillaroscopic findings in microcirculation research. The Cap image analysis system comprises an IBM-compatible PC with a Matrox image processing card and real-time video tape digitalization. The video recorder is driven by a personal computer to which it is connected via an RS-232 interface. In contrast to currently available systems, the program presented here makes it possible to select ana of several integrated image analysis functions, depending on the quality of the video image. Some examples of the analysis functions available are measurements of erythrocyte Bow velocity using the line shift diagram method, the spatial correlation method, and the auto flying spot method. The standard features of the nem program include a number of special functions and automatic movement correction The system thus makes it possible not only to measure numerous morphological parameters such as capillary diameter, length, torquation index and capillary density, but also to perform video densitometric analysis, for example using fluorescent dyes.
引用
收藏
页码:168 / 175
页数:8
相关论文
共 20 条
[1]  
Basler A., 1919, MUNCH MED WSCHR, V13, P347
[2]  
BLOCH HE, 1963, ANGIOLOGY, V14, P97
[3]   RED BLOOD-CELL VELOCITY IN NAILFOLD CAPILLARIES OF MAN MEASURED BY A TELEVISION MICROSCOPY TECHNIQUE [J].
BOLLINGER, A ;
BUTTI, P ;
BARRAS, JP ;
TRACHSLER, H ;
SIEGENTHALER, W .
MICROVASCULAR RESEARCH, 1974, 7 (01) :61-72
[4]  
BRANEMARK P.-I., 1963, BIORHEOLOGY, V1, P143
[5]  
HAHN M, 1995, BRIT J DERMATOL, V133, P704
[6]   Synchronous measurements of blood pressure and red blood cell velocity in capillaries of human skin [J].
Hahn, M ;
Klyscz, T ;
Junger, M .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1996, 106 (06) :1256-1259
[7]   ERROR ANALYSIS, BIOLOGICAL INFLUENCING FACTORS, AND VARIANCE OF PERIUNGUAL VIDEO CAPILLARY MICROSCOPY [J].
JUNG, F ;
TOONEN, H ;
MROWIETZ, C ;
WOLF, S ;
KIESEWETTER, H ;
WENZEL, E ;
GERSONDE, K ;
MULLER, G .
BIOMEDIZINISCHE TECHNIK, 1990, 35 (09) :195-204
[8]  
Jung F, 1992, BIOMED OPTRONICS, V1, P27
[9]  
JUNG F, 1987, BIOMED TECHNIK, V9, P204
[10]  
JUNGER M, 1995, BIOENGINEERING SKIN, P113