Comparison of an ultrasonic nebulizer with a cross-flow nebulizer for selenium speciation by ion-chromatography and inductively coupled plasma mass spectrometry

被引:30
作者
Gammelgaard, B [1 ]
Jons, O [1 ]
机构
[1] Royal Danish Sch Pharm, Dept Analyt & Pharmaceut Chem, DK-2100 Copenhagen, Denmark
关键词
D O I
10.1039/a908975f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The purpose of this work was to compare an ultrasonic nebulizer (USN) with a cross-flow nebulizer (CFN) for selenium speciation with inductively coupled plasma mass spectrometry (ICP-MS) detection. The influence of instrumental parameters as well as composition of the solvent on the selenium species, trimethylselenonium ion (TMSe), selenomethionine (SeMet), selenite and selenate, were studied. Using the CFN, there was no difference in sensitivity of the four selenium species and the species responded similarly to all changes. In the USN, sensitivity increased between 14 and 21 times for the species. Selenate and TMSe showed the largest increase in sensitivity while selenite and SeMet showed lower increases. The species behaved differently on changing the heat temperature of the USN and the pH of the solutions. When the carbon enhancement effect was exploited by adding 3% methanol, the increase in sensitivity with the USN was between 2.5 and 8.7 times. It was observed that yttrium could not be used as internal standard in the USN as the sensitivity of this element was dependent on pH as well as the presence of different selenium species, while rhodium was unaffected by these parameters. The USN was found inapplicable for normal sample uptake as memory effects were severe and low blank values, therefore, difficult to obtain. Thus, this nebulizer could only be used for transient signals. When coupling the nebulizers to an anion-exchange chromatographic system, peak widths of the chromatographic signals were comparable. The sensitivity with the USN was 10 times larger than with the CFN. Detection limits, however, were not decreased correspondingly. Detection limits calculated on the basis of Se-82 were between 0.15 and 0.24 mu g l(-1) Se with the USN, while they were between 0.43 and 1.6 mu g l(-1) Se with the CFN.
引用
收藏
页码:499 / 505
页数:7
相关论文
共 25 条
[1]  
ALAEJOS MS, 1993, CLIN CHEM, V39, P2040
[2]   SIGNAL ENHANCEMENT OF ELEMENTS DUE TO THE PRESENCE OF CARBON-CONTAINING COMPOUNDS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
JAUNAULT, L ;
MAURAS, Y ;
MERMET, JM ;
DELAPORTE, T .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1497-1498
[3]  
BURK RF, 1995, HEPATOLOGY, V21, P561
[4]  
CLARK LC, 1996, JAMA-J AM MED ASSOC, V276, P157
[5]   DIAGNOSTIC STUDIES ON DESOLVATED AEROSOLS FROM ULTRASONIC NEBULIZERS [J].
CLIFFORD, RH ;
SOHAL, P ;
LIU, HY ;
MONTASER, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (09) :1107-1122
[6]   ULTRASONIC NEBULIZATION AND ARSENIC VALENCE STATE CONSIDERATIONS PRIOR TO DETERMINATION VIA INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
CREED, JT ;
MARTIN, TD ;
BROCKHOFF, CA .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (06) :443-447
[7]   SYNTHESIS OF [TRIMETHYLSELENONIUM-SE-75 IODIDE FROM [SELENOCYSTINE-SE-75 [J].
FOSTER, SJ ;
GANTHER, HE .
ANALYTICAL BIOCHEMISTRY, 1984, 137 (01) :205-209
[8]   Determination of selenium in urine by inductively coupled plasma mass spectrometry: interferences and optimization [J].
Gammelgaard, B ;
Jons, O .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (05) :867-874
[9]   ELIMINATION OF INTERFERENCES IN THE DETERMINATION OF ARSENIC AND SELENIUM IN BIOLOGICAL SAMPLES BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
GOOSSENS, J ;
VANHAECKE, F ;
MOENS, L ;
DAMS, R .
ANALYTICA CHIMICA ACTA, 1993, 280 (01) :137-143
[10]   Speciation of arsenic and selenium compounds by HPLC hyphenated to specific detectors: a review of the main separation techniques [J].
Guerin, T ;
Astruc, A ;
Astruc, M .
TALANTA, 1999, 50 (01) :1-24