Custom stems for femoral deformity in patients less than 40 years of age 70 hips followed for an average of 14 years

被引:25
作者
Akbar, Michael [1 ]
Aldinger, Guenther [2 ]
Krahmer, Knut [2 ]
Bruckner, Thomas [3 ]
Aldinger, Peter R. [1 ]
机构
[1] Univ Heidelberg, Dept Orthoped Surg, D-6900 Heidelberg, Germany
[2] Diakonieklinikum, Paulinenhilfe Orthoped Ctr, Stuttgart, Germany
[3] Univ Heidelberg, Dept Med Biometry & Informat, Heidelberg, Germany
关键词
CEMENTING TECHNIQUES; YOUNG-PATIENTS; ARTHROPLASTY; REPLACEMENT; 10-YEAR; PROSTHESIS; COMPONENTS; FIXATION; IMPLANTS; DESIGN;
D O I
10.3109/17453670903062470
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Background and purpose Femoral deformity associated with osteoarthritis is a challenge for both the surgeon and the implant. Many of the patients with these deformities are young. Standard implants can be difficult to fit into these femurs. We prospectively evaluated the outcome of custom uncemented femoral stems in young patients. Methods 61 consecutive patients (72 hips) underwent surgery for osteoarthritis because of femoral deformity at a mean age of 35 (22-40) years. The patients received a CT3D-A custom-made femoral stem and an uncemented cup. The mean follow-up time was 14 (10-16) years. 2 patients died at 7 and 8 years after surgery, otherwise, none of the patients were lost to follow-up. Results At follow-up, the femoral prosthesis had not been revised in 59 patients (70 hips). 3 patients (3 hips) had required revision surgery due to loosening of the acetabular component; 2 hips were awaiting revision surgery for loosening of the acetabular cup. There were no cases of dislocation or infection. At review, all stems were considered stable according to the radiographic criteria. No migration or subsidence was observed on plain radiographs. Interpretation Our results are comparable to published results of custom stems regarding survival and outcome. Considering the young age and the deformities in this series of uncemented custom femoral stems, and the fact that there was follow-up of up to 16 years, the survival is remarkable. This technique appears to be a reasonable alternative in younger patients with femoral deformities.
引用
收藏
页码:420 / 425
页数:6
相关论文
共 25 条
[1]
Changes in proximal femoral strain after insertion of uncemented standard and customised femoral stems [J].
Aamodt, A ;
Lund-Larsen, J ;
Eine, J ;
Andersen, E ;
Benum, P ;
Husby, OS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (06) :921-929
[2]
COMPUTER-AIDED MANUFACTURE OF INDIVIDUAL ENDOPROSTHESES - PRELIMINARY COMMUNICATION [J].
ALDINGER, G ;
FISCHER, A ;
KURTZ, B .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1983, 102 (01) :31-35
[3]
Aldinger PR, 2003, ACTA ORTHOP SCAND, V74, P253, DOI 10.1080/00016470310014157
[4]
BALLARD WT, 1994, J BONE JOINT SURG AM, V76A, P959
[5]
BARGAR WL, 1989, CLIN ORTHOP RELAT R, P73
[6]
Berry DJ, 1999, CLIN ORTHOP RELAT R, P262
[7]
Custom total hip arthroplasty [J].
Bert, JM .
JOURNAL OF ARTHROPLASTY, 1996, 11 (08) :905-915
[8]
ECTOPIC OSSIFICATION FOLLOWING TOTAL HIP-REPLACEMENT - INCIDENCE AND A METHOD OF CLASSIFICATION [J].
BROOKER, AF ;
BOWERMAN, JW ;
ROBINSON, RA ;
RILEY, LH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1973, A 55 (08) :1629-1632
[9]
Ten-year results with hydroxyapatite-coated total hip femoral components in patients less than fifty years old - A concise follow-up of a previous report [J].
Capello, WN ;
D'Antonio, JA ;
Feinberg, JR ;
Manley, MT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (05) :885-889
[10]
Three-dimensional polyethylene wear of a press-fit titanium prosthesis - Factors influencing generation of polyethylene debris [J].
Devane, PA ;
Horne, JG ;
Martin, K ;
Coldham, G ;
Krause, B .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :256-266