A cell-based bar code reader for high-throughput screening of ion channel-ligand interactions

被引:57
作者
Sinclair, J
Pihl, J
Olofsson, J
Karlsson, M
Jardemark, K
Chiu, DT
Orwar, O [1 ]
机构
[1] Chalmers Univ Technol, Dept Chem Phys, SE-41296 Gothenburg, Sweden
[2] Chalmers Univ Technol, Microtechnol Ctr, SE-41296 Gothenburg, Sweden
[3] Univ Washington, Dept Chem, Seattle, WA 98195 USA
[4] Karolinska Inst, Dept Physiol & Pharmacol, Stockholm, Sweden
关键词
D O I
10.1021/ac026133f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper presents a microfluidics-patch clamp platform for performing high-throughput screening and rapid characterization of weak-affinity ion channel-ligand interactions. This platform integrates a microfluidic chip consisting of multiple channels entering an open volume with standard patch clamp equipment. The microfluidic chip is placed on a motorized scanning stage and the method relies on the ability to scan rapidly, on the order of milliseconds, a patch-clamped cell across discrete zones of different solutions created in the open volume. Under ideal conditions, this method has the capacity to obtain kinetically resolved patch clamp measurements and dose-response curves of up to 10(3) ligand solutions in a single day.
引用
收藏
页码:6133 / 6138
页数:6
相关论文
共 24 条
[1]  
Batchelor G.K., 1970, INTRO FLUID DYNAMICS
[2]   Generation of gradients having complex shapes using microfluidic networks [J].
Dertinger, SKW ;
Chiu, DT ;
Jeon, NL ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 2001, 73 (06) :1240-1246
[3]   Screening of ion channel receptor agonists using capillary electrophoresis-patch clamp detection with resensitized detector cells [J].
Farre, C ;
Sjöberg, A ;
Jardemark, K ;
Jacobson, I ;
Orwar, O .
ANALYTICAL CHEMISTRY, 2001, 73 (06) :1228-1233
[4]   Whole cell patch clamp recording performed on a planar glass chip [J].
Fertig, N ;
Blick, RH ;
Behrends, JC .
BIOPHYSICAL JOURNAL, 2002, 82 (06) :3056-3062
[5]   ACTIVATION AND DESENSITIZATION OF EMBRYONIC-LIKE RECEPTOR CHANNELS IN MOUSE MUSCLE BY ACETYLCHOLINE CONCENTRATION STEPS [J].
FRANKE, C ;
KOLTGEN, D ;
HATT, H ;
DUDEL, J .
JOURNAL OF PHYSIOLOGY-LONDON, 1992, 451 :145-158
[6]   IMPROVED PATCH-CLAMP TECHNIQUES FOR HIGH-RESOLUTION CURRENT RECORDING FROM CELLS AND CELL-FREE MEMBRANE PATCHES [J].
HAMILL, OP ;
MARTY, A ;
NEHER, E ;
SAKMANN, B ;
SIGWORTH, FJ .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1981, 391 (02) :85-100
[7]   A rapid diffusion immunoassay in a T-sensor [J].
Hatch, A ;
Kamholz, AE ;
Hawkins, KR ;
Munson, MS ;
Schilling, EA ;
Weigl, BH ;
Yager, P .
NATURE BIOTECHNOLOGY, 2001, 19 (05) :461-465
[8]   PHOSPHORYLATION OF THE NICOTINIC ACETYLCHOLINE-RECEPTOR REGULATES ITS RATE OF DESENSITIZATION [J].
HUGANIR, RL ;
DELCOUR, AH ;
GREENGARD, P ;
HESS, GP .
NATURE, 1986, 321 (6072) :774-776
[9]   The impact of receptor desensitization on fast synaptic transmission [J].
Jones, MV ;
Westbrook, GL .
TRENDS IN NEUROSCIENCES, 1996, 19 (03) :96-101
[10]   Microfabrication inside capillaries using multiphase laminar flow patterning [J].
Kenis, PJA ;
Ismagilov, RF ;
Whitesides, GM .
SCIENCE, 1999, 285 (5424) :83-85