Metal ion analysis in contaminated water samples using anodic stripping voltammetry and a nanocrystalline diamond thin-film electrode

被引:108
作者
Sonthalia, P [1 ]
McGaw, E [1 ]
Show, Y [1 ]
Swain, GM [1 ]
机构
[1] Michigan State Univ, Dept Chem, E Lansing, MI 48824 USA
基金
美国国家航空航天局;
关键词
nanocrystalline diamond electrodes; trace metal ion analysis; anodic stripping voltammetry;
D O I
10.1016/j.aca.2004.06.071
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Boron-doped nanocrystalline diamond thin-film electrodes were employed for the detection and quantification of Ag (I), Cu (II), Pb (II), Cd (II), and Zn (II) in several contaminated water samples using anodic stripping voltammetric (ASV). Diamond is an alternate electrode that possesses many of the same attributes as Hg and, therefore, appears to be a viable material for this electroanalytical measurement. The nanocrystalline form has been found to perform slightly better than the more conventional microcrystalline form of diamond in this application. Differential pulse voltammetry (DPASV) was used to detect these metal ions in lake water, well water, tap water, wastewater treatment sludge, and soil. The electrochemical results were compared with data from inductively coupled plasma mass spectrometric (ICP-MS) and or atomic absorption spectrometric (AAS) measurements of the same samples. Diamond is shown to function well in this electroanalytical application, providing a wide linear dynamic range, a low limit of quantitation, excellent response precision, and good response accuracy. For the analysis of Pb (II), bare diamond provided a response nearly identical to that obtained with a Hg-coated glassy carbon electrode. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:35 / 44
页数:10
相关论文
共 30 条
[1]   ELECTRODEPOSITION OF METAL ADLAYERS ON BORON-DOPED DIAMOND THIN-FILM ELECTRODES [J].
AWADA, M ;
STROJEK, JW ;
SWAIN, GM .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1995, 142 (03) :L42-L45
[2]   Electrochemical detection of trace concentrations of cadmium and lead with a boron-doped diamond electrode:: Effect of KCl and KNO3 electrolytes, interferences and measurement in river water [J].
Babyak, C ;
Smart, RR .
ELECTROANALYSIS, 2004, 16 (03) :175-182
[3]   Determination of nanomolar concentrations of lead and cadmium by anodic-stripping voltammetry at the silver electrode [J].
Bonfil, Y ;
Kirowa-Eisner, E .
ANALYTICA CHIMICA ACTA, 2002, 457 (02) :285-296
[4]  
Brainina KZ, 1997, NATO ASI 2, V38, P255
[5]   Toxicity testing of wastewater and sewage sludge by biosensors, bioassays and chemical analysis [J].
Farré, M ;
Barceló, D .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2003, 22 (05) :299-310
[6]   On-site analysis of arsenic in groundwater using a microfabricated gold ultramicroelectrode array [J].
Feeney, R ;
Kounaves, SP .
ANALYTICAL CHEMISTRY, 2000, 72 (10) :2222-2228
[7]   Standard electrochemical behavior of high-quality, boron-doped polycrystalline diamond thin-film electrodes [J].
Granger, MC ;
Witek, M ;
Xu, JS ;
Wang, J ;
Hupert, M ;
Hanks, A ;
Koppang, MD ;
Butler, JE ;
Lucazeau, G ;
Mermoux, M ;
Strojek, JW ;
Swain, GM .
ANALYTICAL CHEMISTRY, 2000, 72 (16) :3793-3804
[8]   Polycrystalline diamond electrodes: basic properties and applications as amperometric detectors in flow injection analysis and liquid chromatography [J].
Granger, MC ;
Xu, JS ;
Strojek, JW ;
Swain, GM .
ANALYTICA CHIMICA ACTA, 1999, 397 (1-3) :145-161
[9]  
HOLT KB, IN PRESS J PHYS CH B
[10]   An introduction to bismuth film electrode for use in cathodic electrochemical detection [J].
Hutton, EA ;
Ogorevc, B ;
Hocevar, SB ;
Weldon, F ;
Smyth, MR ;
Wang, J .
ELECTROCHEMISTRY COMMUNICATIONS, 2001, 3 (12) :707-711