Nitric oxide-releasing xerogel-based fiber-optic pH sensors

被引:15
作者
Dobmeier, Kevin P. [1 ]
Charville, Gregory W. [1 ]
Schoenfisch, Mark H. [1 ]
机构
[1] Univ N Carolina, Dept Chem, Chapel Hill, NC 27599 USA
关键词
D O I
10.1021/ac060995p
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A xerogel-based optical pH sensor capable of releasing low levels of nitric oxide (NO) and measuring changes in solution pH is reported. Through simple dip-coating procedures, aminoalkoxysilane-based xerogel films modified with N-diazeniumdiolate NO donor precursors and the fluorescent pH indicator seminaphthorhodamine-1 carboxylate (SNARF-1) were sequentially deposited onto optical fibers. The resulting sensors were characterized by fast and linear response to pH throughout the physiological range (pH 7.0-7.8). Real-time chemiluminescence measurements confirmed that the presence of the overlying SNARF-1-containing TMOS layer did not have an inhibitory effect on N-diazeniumdiolate formation or NO release, and the NO-releasing coatings were capable of maintaining NO fluxes > 0.4 pmol/cm(2) s up to 16 h. In vitro blood compatibility studies using porcine platelets confirmed the expected thromboresistivity of the NO-releasing xerogel coatings.
引用
收藏
页码:7461 / 7466
页数:6
相关论文
共 50 条
[1]  
BRINKER J, 1989, SOL GEL SCI
[2]   Stability of new optical pH sensing material based on cross-linked poly(vinyl alcohol) copolymer [J].
Cajlakovic, M ;
Lobnik, A ;
Werner, T .
ANALYTICA CHIMICA ACTA, 2002, 455 (02) :207-213
[3]  
CAZENAVE J, 1993, ROLE PLATELETS BLOOD, P69
[4]   Preparation and characterization of implantable sensors with nitric oxide release coatings [J].
Frost, MC ;
Batchelor, MM ;
Lee, YM ;
Zhang, HP ;
Kang, YJ ;
Oh, BK ;
Wilson, GS ;
Gifford, R ;
Rudich, SM ;
Meyerhoff, ME .
MICROCHEMICAL JOURNAL, 2003, 74 (03) :277-288
[5]   In vivo biocompatibility and analytical performance of intravascular amperometric oxygen sensors prepared with improved nitric oxide-releasing silicone rubber coating [J].
Frost, MC ;
Rudich, SM ;
Zhang, HP ;
Maraschio, MA ;
Meyerhoff, ME .
ANALYTICAL CHEMISTRY, 2002, 74 (23) :5942-5947
[6]   Implantable chemical sensors for real-time clinical monitoring: progress and challenges [J].
Frost, MC ;
Meyerhoff, ME .
CURRENT OPINION IN CHEMICAL BIOLOGY, 2002, 6 (05) :633-641
[7]   Continuous intravascular blood gas monitoring: development, current techniques, and clinical use of a commercial device [J].
Ganter, M ;
Zollinger, A .
BRITISH JOURNAL OF ANAESTHESIA, 2003, 91 (03) :397-407
[8]   Accuracy and performance of a modified continuous intravascular blood gas monitoring device during thoracoscopic surgery [J].
Ganter, MT ;
Hofer, CK ;
Zollinger, A ;
Spahr, T ;
Pasch, T ;
Zalunardo, MP .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2004, 18 (05) :587-591
[9]   Mediation of in vivo glucose sensor inflammatory response via nitric oxide release [J].
Gifford, R ;
Batchelor, MM ;
Lee, Y ;
Gokulrangan, G ;
Meyerhoff, ME ;
Wilson, GS .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2005, 75A (04) :755-766
[10]   AN EVANESCENT-WAVE BIOSENSOR .2. FLUORESCENT SIGNAL ACQUISITION FROM TAPERED FIBER OPTIC PROBES [J].
GOLDEN, JP ;
ANDERSON, GP ;
RABBANY, SY ;
LIGLER, FS .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (06) :585-591