A new modular chemiluminescence immunoassay analyser evaluated

被引:24
作者
Ognibene, A
Drake, CJ
Jeng, KYS
Pascucci, TE
Hsu, S
Luceri, F
Messeri, G
机构
[1] Azienda Osped Careggi, Lab Endocrinol & Toxicol, Florence, Italy
[2] Abbott Labs, Abbott Pk, IL 60064 USA
关键词
immunoassay; automation; chemiluminescence; evaluation;
D O I
10.1515/CCLM.2000.037
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Thyrotropin (TSH), free thyroxine (fT4) and testosterone assays have been used as a probe to evaluate the performances of a new modular chemiluminescence (CL) immunoassay analyser, the Abbott Architect(TM) I2000. The evaluation was run in parallel on other systems that use CL as the detection reaction: DPC Immulite(R), Chiron Diagnostics ACS-180 and ACS Centaur (TSH functional sensitivity only). TSH functional sensitivity was 0.0012, 0.009, 0.033 and 0.039 mU/I for the Architect, Immulite, ACS Centaur and ACS-180, respectively. Testosterone functional sensitivity was 0.38, 3.7 and 2.0 nmol/l for Architect, Immulite and ACS-180, respectively. Good correlation was obtained between the ACS-180 and Architect for all assays. The Immulite correlation did not agree well with the Architect or ACS-180 for fT4 and testosterone but was in good agreement for TSH. Regarding fT4 and testosterone, equilibrium dialysis and isotopic dilution gas-chromatography mass-spectrometry IGC-MS) respectively were used as reference methods. For both within- and between-run precision, the Architect showed the best reproducibility for all three analytes (CV < 6%).
引用
收藏
页码:251 / 260
页数:10
相关论文
共 11 条
[1]   Letter: Detection of reaction intermediates by flow injection electrospray ionization mass spectrometry: reaction of chemiluminescent N-sulfonylacridinium-9-carboxamides with hydrogen peroxide [J].
Adamczyk, M ;
Fishpaugh, JR ;
Gebler, JC ;
Mattingly, PG ;
Shreder, K .
EUROPEAN MASS SPECTROMETRY, 1998, 4 (02) :121-125
[2]   Neopentyl 3-triflyloxypropanesulfonate. A reactive sulfopropylation reagent for the preparation of chemiluminescent labels [J].
Adamczyk, M ;
Chen, YY ;
Mattingly, PG ;
Pan, Y ;
Rege, S .
JOURNAL OF ORGANIC CHEMISTRY, 1998, 63 (16) :5636-5639
[3]  
BABSON AL, 1991, J CLIN IMMUNOASSAY, V14, P83
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
DUDLEY RF, 1991, J CLIN IMMUNOASSAY, V14, P77
[6]  
MALAVASI B, 1992, EUR J CLIN CHEM CLIN, V30, P157
[7]   DETERMINATION OF TESTOSTERONE AND ITS TISSUE METABOLITES (DHT AND 3-ALPHA-DIOL) IN HUMAN-PLASMA AND PROSTATIC TISSUE BY ISOTOPIC DILUTION MASS-SPECTROMETRY [J].
MONETI, G ;
GIOVANNINI, MG ;
GUARNA, A ;
FORTI, G ;
SALERNO, R ;
MAGINI, A ;
SERIO, M .
JOURNAL OF STEROID BIOCHEMISTRY AND MOLECULAR BIOLOGY, 1987, 27 (1-3) :53-59
[8]  
NELSON JC, 1988, CLIN CHEM, V34, P1737
[9]  
PASSING H, 1983, J CLIN CHEM CLIN BIO, V21, P709
[10]  
PRICE A, 1995, CLIN CHEM ENZYM COMM, V6, P373