A new modular chemiluminescence immunoassay analyser evaluated

被引:24
作者
Ognibene, A
Drake, CJ
Jeng, KYS
Pascucci, TE
Hsu, S
Luceri, F
Messeri, G
机构
[1] Azienda Osped Careggi, Lab Endocrinol & Toxicol, Florence, Italy
[2] Abbott Labs, Abbott Pk, IL 60064 USA
关键词
immunoassay; automation; chemiluminescence; evaluation;
D O I
10.1515/CCLM.2000.037
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Thyrotropin (TSH), free thyroxine (fT4) and testosterone assays have been used as a probe to evaluate the performances of a new modular chemiluminescence (CL) immunoassay analyser, the Abbott Architect(TM) I2000. The evaluation was run in parallel on other systems that use CL as the detection reaction: DPC Immulite(R), Chiron Diagnostics ACS-180 and ACS Centaur (TSH functional sensitivity only). TSH functional sensitivity was 0.0012, 0.009, 0.033 and 0.039 mU/I for the Architect, Immulite, ACS Centaur and ACS-180, respectively. Testosterone functional sensitivity was 0.38, 3.7 and 2.0 nmol/l for Architect, Immulite and ACS-180, respectively. Good correlation was obtained between the ACS-180 and Architect for all assays. The Immulite correlation did not agree well with the Architect or ACS-180 for fT4 and testosterone but was in good agreement for TSH. Regarding fT4 and testosterone, equilibrium dialysis and isotopic dilution gas-chromatography mass-spectrometry IGC-MS) respectively were used as reference methods. For both within- and between-run precision, the Architect showed the best reproducibility for all three analytes (CV < 6%).
引用
收藏
页码:251 / 260
页数:10