Direct selenium x-ray detector for fluoroscopy, R&F, and radiography

被引:59
作者
Choquette, M [1 ]
Rougeot, H [1 ]
Martin, JP [1 ]
Laperrière, L [1 ]
Shukri, Z [1 ]
Polischuk, B [1 ]
机构
[1] ANRAD Corp, St Laurent, PQ H4R 2P1, Canada
来源
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING | 2000年 / 3977卷
关键词
selenium; flat-panel detector; active matrix; TFT; direct; fluoroscopy; real-time; dynamic;
D O I
10.1117/12.384485
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of selenium alloys for direct conversion fluoroscopy flat-panel detectors has been underestimated. The purpose of this paper is to demonstrate the salient features of a selenium-based detector designed for R&F applications. The detector has an active area of 30cm x 27cm and comprises 2048 x 1792 pixels at 150 mu m pitch. The geometric fill factor is 66%, but experimental evidence supports the fact that internal electric field bending leads to an effective fill factor approaching unity. This detector is designed to support full resolution images at 15 frames/second, and 896x1024 resolution at 30 frames/second. The detector is coated with a simple coplanar "p-i-n" selenium diode structure which has a dark current less than 100pA/cm(2) The thickness of this structure is 1000 mu m to absorb 77% of a NEMA standard fluoroscopy beam. Measurements show we have obtained an x-ray sensitivity of 4400pC/mR/cm(2), which translates to 1212 collected charges per absorbed x-ray. Resolution was measured to be near the theoretically predicted values. with a modulation of 63% at the Nyquist limit of 3.331p/mm. Phantom images were obtained at a frame rate of 15frames per second, and negligible lag was observed in this image sequence.
引用
收藏
页码:128 / 136
页数:9
相关论文
共 14 条
[1]   A large-area, 97 μm pitch, indirect-detection, active matrix, flat-panel imager (AMFPI) [J].
Antonuk, LE ;
El-Mohri, Y ;
Hall, A ;
Jee, KW ;
Maolinbay, M ;
Nassif, SC ;
Rong, XJ ;
Siewerdsen, JH ;
Zhao, QH ;
Weisfield, RL .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :2-13
[2]   X-ray imaging using amorphous selenium: Determination of Swank factor by pulse height spectroscopy [J].
Blevis, IM ;
Hunt, DC ;
Rowlands, JA .
MEDICAL PHYSICS, 1998, 25 (05) :638-641
[3]   Technical and clinical results of an experimental Flat Dynamic (digital) X-ray image Detector (FDXD) system with real-time corrections [J].
Bruijns, TJC ;
Alving, PL ;
Baker, EL ;
Bury, R ;
Cowen, AR ;
Jung, N ;
Luijendijk, HA ;
Meulenbrugge, HJ ;
Stouten, HJ .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :33-44
[4]   Amorphous silicon x-ray image sensor [J].
Chabbal, J ;
Chaussat, C ;
Ducourant, T ;
Fritsch, L ;
Michailos, J ;
Spinnler, V ;
Vieux, G ;
Arques, M ;
Hahm, G ;
Hoheisel, M ;
Horbaschek, H ;
Schulz, R ;
Spahn, M .
PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 :499-510
[5]   Characterization of an amorphous silicon fluoroscopic imager [J].
Colbeth, RE ;
Allen, MJ ;
Day, DJ ;
Gilblom, DL ;
KlausmeierBrown, ME ;
Pavkovich, J ;
Seppi, EJ ;
Shapiro, EG .
PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 :42-51
[6]   A SIMPLE METHOD FOR DETERMINING THE MODULATION TRANSFER-FUNCTION IN DIGITAL RADIOGRAPHY [J].
FUJITA, H ;
TSAI, DY ;
ITOH, T ;
DOI, K ;
MORISHITA, J ;
UEDA, K ;
OHTSUKA, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (01) :34-39
[7]   Detective quantum efficiency of direct, flat panel x-ray imaging detectors for fluoroscopy [J].
Hunt, DC ;
Zhao, W ;
Rowlands, JA .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :408-417
[8]  
LEE DL, 1995, P SOC PHOTO-OPT INS, V2432, P237, DOI 10.1117/12.208342
[9]  
MURPHY G, 1993, SPIE PHYSICS MED IMA, V1896, P174
[10]   A new large area X-ray image sensor [J].
Ouimette, DR ;
Nudelman, S ;
Aikens, R .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :470-476