Ability of calibration phantom to reduce the interscan variability in electron beam computed tomography

被引:10
作者
Budoff, MJ
Mao, SS
Lu, B
Takasu, J
Child, J
Carson, S
Fisher, H
机构
[1] Harbor UCLA Res & Educ Inst, Div Cardiol, Torrance, CA 90502 USA
[2] Harbor UCLA Res & Educ Inst, Div Radiol, Torrance, CA 90502 USA
[3] XiJing Hosp, Dept Radiol, Xian, Peoples R China
关键词
computed tomography; electron beam; coronary artery calcification; reproducibility; calibration phantom;
D O I
10.1097/00004728-200211000-00005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To test the hypothesis that a calibration phantom would improve interpatient and interscan variability in coronary artery calcium (CAC) studies. Methods: We scanned 144 patients twice with or without the calibration phantom and then scanned 93 patients with a single calcific lesion twice and, finally, scanned a cork heart with calcific foci. Results: There were no linear correlations in computed tomography Hounsfield unit (CT HU) and CT HU interscan variation between blood pool and phantom plug,,, at any slice level in patient groups (p > 0.05). The CT HU interscan variation in phantom plugs (2.11 HU) was less than that of the blood pool (3,47 HU p < 0.05) and CAC lesion (20.39; p < 0.001). Comparing images with and without it calibration phantom, there was a significant decrease in CT HU as well as an increase in noise and peak values in patient studies and the cork phantom study. Conclusion: The CT HU attenuation variations of the interpatient and interscan blood pool, calibration phantom plug, and cork coronary arteries were not parallel, Therefore, the ability to adjust the CT HU variation of calcific lesions by a calibration phantom is problematic and may worsen the problem.
引用
收藏
页码:886 / 891
页数:6
相关论文
共 19 条
[1]   QUANTIFICATION OF CORONARY-ARTERY CALCIUM USING ULTRAFAST COMPUTED-TOMOGRAPHY [J].
AGATSTON, AS ;
JANOWITZ, WR ;
HILDNER, FJ ;
ZUSMER, NR ;
VIAMONTE, M ;
DETRANO, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) :827-832
[2]   Preview method for electron-beam CT scanning of the coronary arteries [J].
Bakhsheshi, H ;
Mao, SS ;
Budoff, MJ ;
Bin, L ;
Brundage, BH .
ACADEMIC RADIOLOGY, 2000, 7 (08) :620-626
[3]   Electron beam computed tomography: Screening for coronary artery disease [J].
Budoff, MJ ;
Brundage, BH .
CLINICAL CARDIOLOGY, 1999, 22 (09) :554-558
[4]  
Budoff MJ, 2001, AM J CARDIOL, V88, p46E
[5]   Electron-beam CT: The effect of using a correction function on coronary artery calcium quantitation [J].
Greaser, LE ;
Yoon, HC ;
Mather, RT ;
McNitt-Gray, M ;
Goldin, JG .
ACADEMIC RADIOLOGY, 1999, 6 (01) :40-48
[6]   Improving coronary heart disease risk assessment in asymptomatic people - Role of traditional risk factors and noninvasive cardiovascular tests [J].
Greenland, P ;
Smith, SC ;
Grundy, SM .
CIRCULATION, 2001, 104 (15) :1863-1867
[7]   COMPUTERIZED TOMOGRAPHIC DENSITOMETRY OF THE SOLITARY PULMONARY NODULE USING A NODULE PHANTOM [J].
JONES, FA ;
WIEDEMANN, HP ;
ODONOVAN, PB ;
STOLLER, JK .
CHEST, 1989, 96 (04) :779-784
[8]   THE EUROPEAN SPINE PHANTOM - A TOOL FOR STANDARDIZATION AND QUALITY-CONTROL IN SPINAL BONE-MINERAL MEASUREMENTS BY DXA AND QCT [J].
KALENDER, WA ;
FELSENBERG, D ;
GENANT, HK ;
FISCHER, M ;
DEQUEKER, J ;
REEVE, J .
EUROPEAN JOURNAL OF RADIOLOGY, 1995, 20 (02) :83-92
[9]   CT densitometry of the lungs: Scanner performance [J].
Kemerink, GJ ;
Lamers, RJS ;
Thelissen, GRP ;
vanEngelshoven, JMA .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (01) :24-33
[10]   THE UNRELIABILITY OF CT-NUMBERS AS ABSOLUTE VALUES [J].
LEVI, C ;
GRAY, JE ;
MCCULLOUGH, EC ;
HATTERY, RR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1982, 139 (03) :443-447