Characterization of a fluoroscopic imaging system for kV and MV radiography

被引:20
作者
Drake, DG
Jaffray, DA
Wong, JW
机构
[1] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI 48073 USA
[2] Oakland Univ, Dept Phys, Rochester, MI 48309 USA
关键词
megavoltage; kilovoltage; detective quantum efficiency; portal imaging;
D O I
10.1118/1.598955
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An on-line kilovoltage (kV) imaging system has been implemented on a medical linear accelerator to verify radiotherapy field placement. A kV x-ray tube is mounted on the accelerator at 90 degrees to the megavoltage (MV) source and shares the same isocenter. Nearly identical CCD-based fluoroscopic imagers are mounted opposite the two x-ray sources. These systems are being used in a clinical study of patient setup error that examines the advantage of kV imaging for on-line localization. In the investigation reported here, the imaging performance of the kV and MV systems are characterized to provide support to the conclusions of the studies of setup error. A spatial-frequency-dependent linear systems model is used to predict the detective quantum efficiencies (DQEs) of the two systems. Each is divided into a series of gain and spreading stages. The parameters of each stage are either measured or obtained from the literature. The model predicts the system gain to within 7% of the measured gain for the MV system and to within 10% for the kV system. The systems' noise power spectra (NPSs) and modulation transfer functions (MTFs) are measured to construct the measured DQEs. X-ray fluences are calculated using modeled polyenergetic spectra. Measured DQEs agree well with those predicted by the model. The model reveals that the MV system is well optimized, and is x-ray quantum noise limited at low spatial frequencies. The kV system is suboptimal, but for purposes of patient positioning yields images superior to those produced by the MV system. This is attributed to the kV system's higher DQE and to the inherently higher contrasts present at kV energies. (C) 2000 American Association of Physicists in Medicine.
引用
收藏
页码:898 / 905
页数:8
相关论文
共 36 条
  • [1] Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imaging
    Antonuk, LE
    El-Mohri, Y
    Huang, WD
    Jee, KW
    Siewerdsen, JH
    Maolinbay, M
    Scarpine, VE
    Sandler, H
    Yorkston, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (02): : 437 - 454
  • [2] A quantum accounting and detective quantum efficiency analysis for video-based portal imaging
    Bissonnette, JP
    Cunningham, IA
    Jaffray, DA
    Fenster, A
    Munro, P
    [J]. MEDICAL PHYSICS, 1997, 24 (06) : 815 - 826
  • [3] Optimal phosphor thickness for portal imaging
    Bissonnette, JP
    Cunningham, IA
    Munro, P
    [J]. MEDICAL PHYSICS, 1997, 24 (06) : 803 - 814
  • [4] STUDIES OF X-RAY-ENERGY ABSORPTION AND QUANTUM NOISE PROPERTIES OF X-RAY SCREENS BY USE OF MONTE-CARLO SIMULATION
    CHAN, HP
    DOI, K
    [J]. MEDICAL PHYSICS, 1984, 11 (01) : 37 - 46
  • [5] A SPATIAL-FREQUENCY DEPENDENT QUANTUM ACCOUNTING DIAGRAM AND DETECTIVE QUANTUM EFFICIENCY MODEL OF SIGNAL AND NOISE-PROPAGATION IN CASCADED IMAGING-SYSTEMS
    CUNNINGHAM, IA
    WESTMORE, MS
    FENSTER, A
    [J]. MEDICAL PHYSICS, 1994, 21 (03) : 417 - 427
  • [6] Degradation of the detective quantum efficiency due to a non-unity detector fill factor
    Cunningham, IA
    [J]. PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 : 22 - 31
  • [7] CUNNINGHAM IA, 1997, EXPANDING ROLE MED P, P231
  • [8] DOI K, 1982, MTFS WEINER SPECTRA
  • [9] A prototype amorphous silicon array based radiotherapy portal imager
    Drake, DG
    Jaffray, DA
    Wong, JW
    [J]. PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 : 32 - 41
  • [10] GALBRAITH DM, 1989, PHYS MED BIOL, V34, P734