Luminescence lifetime imaging of oxygen, pH, and carbon dioxide distribution using optical sensors

被引:169
作者
Liebsch, G
Klimant, I [1 ]
Frank, B
Holst, G
Wolfbeis, OS
机构
[1] Univ Regensburg, Inst Analyt Chemo Biosensors, D-93040 Regensburg, Germany
[2] Max Planck Inst Marine Microbiol, D-28359 Bremen, Germany
关键词
optical sensor films; time-resolved imaging system; microsecond decay time sensors;
D O I
10.1366/0003702001949726
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We present a modular system for time-resolved two-dimensional luminescence lifetime imaging of planar optical chemical sensors. It is based on a fast, gateable charge-coupled device (CCD) camera without image intensifier and a pulsable light-emitting diode (LED) array as a light source. Software was developed for data acquisition with a maximum of parameter variability and for background suppression. This approach allows the operation of the system even under daylight. Optical sensors showing analyte-specific changes of their luminescence decay time were tested and used for sensing pO(2), pCO(2), pH, and temperature. The luminophores employed are either platinum(II)-porphyrins or ruthenium(II)-polypyridyl completes, contained in polymer films, and can be efficiently excited by blue LEDs. The decay times of the sensor films vary from 70 mu s far the Pt(II)-porphyrins to several 100 ns for the Ru(II) complexes. In a typical application, 7 mm-diameter spots of the respective optical sensor films were placed at the bottom of the wells of microtiterplates. Thus, every well represents a separate calibration chamber with an integrated sensor element. Both luminescence intensity-based and time-resolved images of the sensor spots were evaluated and compared. The combination of optical sensor technology with time-resolved imaging allows a determination of the distribution of chemical or physical parameters in heterogeneous systems and is therefore a powerful tool for screening and mapping applications.
引用
收藏
页码:548 / 559
页数:12
相关论文
共 31 条
  • [1] DETERMINATION OF OXYGEN CONCENTRATIONS BY LUMINESCENCE QUENCHING OF A POLYMER-IMMOBILIZED TRANSITION-METAL COMPLEX
    BACON, JR
    DEMAS, JN
    [J]. ANALYTICAL CHEMISTRY, 1987, 59 (23) : 2780 - 2785
  • [2] AN ERROR ANALYSIS OF THE RAPID LIFETIME DETERMINATION METHOD FOR THE EVALUATION OF SINGLE EXPONENTIAL DECAYS
    BALLEW, RM
    DEMAS, JN
    [J]. ANALYTICAL CHEMISTRY, 1989, 61 (01) : 30 - 33
  • [3] A real time system for Fluorescence Lifetime imaging
    Cubeddu, R
    Canti, G
    Pifferi, A
    Taroni, P
    Valentini, G
    [J]. BIOMEDICAL SENSING, IMAGING, AND TRACKING TECHNOLOGIES II, 1997, 2976 : 98 - 104
  • [4] 2-D fluorescence lifetime imaging using a time-gated image intensifier
    Dowling, K
    Hyde, SCW
    Dainty, JC
    French, PMW
    Hares, JD
    [J]. OPTICS COMMUNICATIONS, 1997, 135 (1-3) : 27 - 31
  • [5] FRANCONI S, 1996, EUR J PEDIATR, V155, P1043
  • [6] Gadella T. W. J., 1997, J FLUORESC, V7, P35
  • [7] Planar optrodes: A new tool for fine scale measurements of two-dimensional O-2 distribution in benthic communities
    Glud, RN
    Ramsing, NB
    Gundersen, JK
    Klimant, I
    [J]. MARINE ECOLOGY PROGRESS SERIES, 1996, 140 (1-3) : 217 - 226
  • [8] Oxygen flux fluorescence lifetime imaging
    Hartmann, P
    Ziegler, W
    Holst, G
    Lubbers, DW
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1997, 38 (1-3) : 110 - 115
  • [9] A modular luminescence lifetime imaging system for mapping oxygen distribution in biological samples
    Holst, G
    Kohls, O
    Klimant, I
    König, B
    Kühl, M
    Richter, T
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1998, 51 (1-3) : 163 - 170
  • [10] König K, 1998, CELL MOL BIOL, V44, P763