共 21 条
Cold neutron prompt gamma-ray activation analysis at NIST - Recent developments
被引:58
作者:
Paul, RL
[1
]
Lindstrom, RM
[1
]
Heald, AE
[1
]
机构:
[1] NATL INST STAND & TECHNOL,DIV REACTOR RADIAT,GAITHERSBURG,MD 20899
关键词:
D O I:
10.1007/BF02109879
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
The cold neutron capture prompt gamma-ray activation analysis (CNPGAA) spectrometer located in the Cold Neutron Research Facility (CNRF) at NIST has proven useful for the analysis of hydrogen and other elements in a wide variety of materials. Modifications of the instrument and the CNRF have resulted in improved measurement capabilities for PGAA. The addition of an atmosphere-controlled sample chamber and Compton suppression have reduced gamma-ray background and increased signal-to-noise ratio. More recent revisions are expected to yield still further improvement in analytical capabilities. Replacement of the D2O ice cold source with a liquid H-2 moderator is expected to yield a 5-10 fold increase in neutron capture rate, and improved neutron and gamma-ray shielding will result in further reduction of the background Other modifications to the instrument allow easier sample mounting and mon: precise positioning of samples in the neutron beam. Significant improvements in detection limits and analytical accuracy are expected.
引用
收藏
页码:63 / 68
页数:6
相关论文