Noninvasive Determination of Cardiac Output by the Inert-Gas-Rebreathing Method - Comparison with Cardiovascular Magnetic Resonance Imaging

被引:45
作者
Saur, Joachim [1 ]
Fluechter, Stephan [1 ]
Trinkmann, Frederik [1 ]
Papavassiliu, Theano [1 ]
Schoenberg, Stefan [2 ]
Weissmann, Joerg [1 ]
Haghi, Dariusch [1 ]
Borggrefe, Martin [1 ]
Kaden, Jens J. [1 ]
机构
[1] Heidelberg Univ, Med Fac Mannheim, Dept Med Cardiol Angiol Pneumol Intens Care 1, D-6800 Mannheim, Germany
[2] Heidelberg Univ, Med Fac Mannheim, Inst Clin Radiol & Nucl Med, D-6800 Mannheim, Germany
关键词
Cardiac output; Inert gas; Cardiovascular magnetic resonance imaging; Rebreathing; CRITICALLY-ILL PATIENTS; HEART-FAILURE; CLINICAL-EVALUATION; SURGICAL-PATIENTS; BLOOD-FLOW; THERMODILUTION; VENTILATION; AGREEMENT; CARE; SETTINGS;
D O I
10.1159/000232407
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
Background: An easy, noninvasive and accurate technique for measuring cardiac output (CO) would be desirable for the diagnosis and therapy of cardiac diseases. Innocor, a novel inert-gas-rebreathing (IGR) system, has shown promising results in smaller studies. An extensive evaluation in a larger, less homogeneous patient collective is lacking. Methods: We prospectively assessed the accuracy and reproducibility of CO measurements obtained by IGR in 305 consecutive patients as compared to the noninvasive gold standard, cardiovascular magnetic resonance (CMR) imaging. Results: Bland-Altman analysis showed a good correspondence of the two methods for CO measurement with an average deviation of 0.2 +/- 1.0 liters/min (mean +/- SD) and a good reproducibility with a mean bias of 0.2 +/- 0.5 liters/min. The accuracy of the present measurements at rest was significantly better in the physiological range than in higher or lower CO ranges. The error levels set forth by current recommendations were exceeded. Conclusion: The data show that IGR measurements are easy to perform and show good agreement with CMR; however, the technique appears to be less accurate in extreme CO ranges at rest. The clinical importance of the IGR method remains to be proven by further studies. Copyright (C) 2009 S. Karger AG, Basel
引用
收藏
页码:247 / 254
页数:8
相关论文
共 31 条
[1]
Noninvasive measurement of cardiac output during exercise by inert gas rebreathing technique: A new tool for heart failure evaluation [J].
Agostoni, P ;
Cattadori, G ;
Apostolo, A ;
Contini, M ;
Palermo, P ;
Marenzi, G ;
Wasserman, K .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (09) :1779-1781
[2]
COMPARISON OF MEASUREMENTS OF CARDIAC-OUTPUT BY BIOIMPEDANCE AND THERMODILUTION IN SEVERELY ILL SURGICAL PATIENTS [J].
APPEL, PL ;
KRAM, HB ;
MACKABEE, J ;
FLEMING, AW ;
SHOEMAKER, WC .
CRITICAL CARE MEDICINE, 1986, 14 (11) :933-935
[3]
Comparison of left ventricular ejection fraction and volumes in heart failure by echocardiography, radionuclide ventriculography and cardiovascular magnetic resonance - Are they interchangeable? [J].
Bellenger, NG ;
Burgess, MI ;
Ray, SG ;
Lahiri, A ;
Coats, AJS ;
Cleland, JGF ;
Pennell, DJ .
EUROPEAN HEART JOURNAL, 2000, 21 (16) :1387-1396
[4]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]
Development and validation of a noninvasive method to estimate cardiac output using cuff sphygmomanometry [J].
Chid, Shiu-Shin ;
Tsai, Jeffrey J. ;
Hsu, Yen-Ming ;
Lapointe, Jeffrey C. ;
Huynh-Covey, Thao ;
Kwan, Oi Ling B. ;
DeMaria, Anthony N. .
CLINICAL CARDIOLOGY, 2007, 30 (12) :615-620
[6]
Thermodilution versus inert gas rebreathing for estimation of effective pulmonary blood flow [J].
Christensen, P ;
Clemensen, P ;
Andersen, PK ;
Henneberg, SW .
CRITICAL CARE MEDICINE, 2000, 28 (01) :51-56
[7]
A MODIFIED PHOTOACOUSTIC AND MAGNETOACOUSTIC MULTIGAS ANALYZER APPLIED IN GAS-EXCHANGE MEASUREMENTS [J].
CLEMENSEN, P ;
CHRISTENSEN, P ;
NORSK, P ;
GRONLUND, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1994, 76 (06) :2832-2839
[8]
The effectiveness of right heart catheterization in the initial care of critically ill patients [J].
Connors, AF ;
Speroff, T ;
Dawson, NV ;
Thomas, C ;
Harrell, FE ;
Wagner, D ;
Desbiens, N ;
Goldman, L ;
Wu, AW ;
Califf, RM ;
Fulkerson, WJ ;
Vidaillet, H ;
Broste, S ;
Bellamy, P ;
Lynn, J ;
Knaus, WA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11) :889-897
[9]
A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[10]
Effects of ventilation on cardiac output determined by inert gas rebreathing [J].
Damgaard, M ;
Norsk, P .
CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2005, 25 (03) :142-147