Rapid quantification of bacterial cells in potable water using a simplified microfluidic device

被引:29
作者
Sakamoto, Chieko
Yamaguchi, Nobuyasu
Yamada, Masumi
Nagase, Hiroyasu
Seki, Minoru
Nasu, Masao
机构
[1] Osaka Univ, Grad Sch Pharmaceut Sci, Suita, Osaka 5650871, Japan
[2] Univ Tokyo, Sch Engn, Bunkyo Ku, Tokyo 1138656, Japan
[3] Osaka Prefecture Univ, Grad Sch Engn, Sakai, Osaka 5998531, Japan
关键词
microfluidic device; potable water; total direct count;
D O I
10.1016/j.mimet.2006.11.003
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simplified microfluidic device for quantification of bacteria in potable water was fabricated and examined. Comparisons of counts of Escherichia coli by the microfluidic system and by epifluorescence microscopy closely correlated (r(2)=0.99). Bacteria in natural mineral water and in purified household tap water were accurately enumerated by using this system within 15 min after fluorescent staining. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:643 / 647
页数:5
相关论文
共 30 条
[1]   Modular concept of a laboratory on a chip for chemical and biochemical analysis [J].
Blankenstein, G ;
Larsen, UD .
BIOSENSORS & BIOELECTRONICS, 1998, 13 (3-4) :427-438
[2]  
Cabrera CR, 2001, ELECTROPHORESIS, V22, P355, DOI 10.1002/1522-2683(200101)22:2<355::AID-ELPS355>3.0.CO
[3]  
2-C
[4]   Flow cytometry and cell sorting of heterogeneous microbial populations: The importance of single-cell analyses [J].
Davey, HM ;
Kell, DB .
MICROBIOLOGICAL REVIEWS, 1996, 60 (04) :641-+
[5]   Total counts, culturable and viable, and non-culturable microflora of a French mineral water:: a case study [J].
Defives, C ;
Guyard, S ;
Oularé, MM ;
Mary, P ;
Hornez, JP .
JOURNAL OF APPLIED MICROBIOLOGY, 1999, 86 (06) :1033-1038
[6]   Rapid prototyping of microfluidic systems in poly(dimethylsiloxane) [J].
Duffy, DC ;
McDonald, JC ;
Schueller, OJA ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :4974-4984
[7]   An integrated microfabricated cell sorter [J].
Fu, AY ;
Chou, HP ;
Spence, C ;
Arnold, FH ;
Quake, SR .
ANALYTICAL CHEMISTRY, 2002, 74 (11) :2451-2457
[8]   Micromachined impedance spectroscopy flow cytometer for cell analysis and particle sizing [J].
Gawad, S ;
Schild, L ;
Renaud, P .
LAB ON A CHIP, 2001, 1 (01) :76-82
[9]   USE OF NUCLEPORE FILTERS FOR COUNTING BACTERIA BY FLUORESCENCE MICROSCOPY [J].
HOBBIE, JE ;
DALEY, RJ ;
JASPER, S .
APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 1977, 33 (05) :1225-1228
[10]  
Hong JW, 2001, ELECTROPHORESIS, V22, P328