Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner

被引:90
作者
Livingston, BJ [1 ]
Chmell, MJ [1 ]
Spector, M [1 ]
Poss, R [1 ]
机构
[1] ROCKFORD ORTHOPED ASSOCIATES, ROCKFORD, IL 61108 USA
关键词
D O I
10.2106/00004623-199710000-00010
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We observed early failure and radiographic signs of accelerated wear as early as one to three years after insertion, without cement, of a metal-backed acetabular component with a liner made of Hylamer (DePuy), This finding prompted us to review a larger cohort of patients in whom that liner had been used, Two hundred and thirty-three components with a Hylamer liner, evaluated at a minimum of two years, had a mean rate of wear of 0.27 millimeter per year compared with 0.12 millimeter per year for a contemporaneous group of fifty acetabular cups with a conventional ultra-high molecular weight polyethylene liner made by another manufacturer, At a mean of 3.2 years, we found a significant difference (p < 0.000000006) between the mean rate of wear (0.20 millimeter per year) when the Hylamer liner articulated with a DePuy modular cobalt-chromium femoral head and the mean rate (0.29 millimeter per year) when the liner articulated with an Osteonics modular cobalt-chromium femoral head, Radiographic evaluation revealed a significant correlation between the total linear wear and the prevalence of osteolytic lesions (r(2) = 0.76, linear regression analysis), We found that wear of 1.5 millimeters or more could be detected by the unaided eye, Because of the positive correlation between osteolysis and wear of 1.5 millimeters or more, we defined a hip with a liner that had that amount of wear as a hip at risk, We concluded that the wear characteristics of a Hylamer liner in vivo are inferior to those of a conventional ultra-high molecular weight polyethylene liner, Also, the rate of wear of the liner is greater when the femoral head is from a manufacturer other than DePuy. A patient who has a total hip replacement that includes a Hylamer liner should be monitored frequently for signs of wear and osteolytic changes, Additional investigations, with longer durations of follow-up and larger populations, are needed to understand fully the importance of our findings.
引用
收藏
页码:1529 / 1538
页数:10
相关论文
共 24 条
[1]  
[Anonymous], 1979, CLIN ORTHOP RELAT RE
[2]  
[Anonymous], CLIN ORTHOPAEDICS
[3]   THE EFFECT OF CONFORMITY AND PLASTIC THICKNESS ON CONTACT STRESSES IN METAL-BACKED PLASTIC IMPLANTS [J].
BARTEL, DL ;
BURSTEIN, AH ;
TODA, MD ;
EDWARDS, DL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1985, 107 (03) :193-199
[4]  
CALLAGHAN JJ, 1995, CLIN ORTHOP RELAT R, V317, P14
[5]   Early failure of Hylamer acetabular inserts due to eccentric wear [J].
Chmell, MJ ;
Poss, R ;
Thomas, WH ;
Sledge, CB .
JOURNAL OF ARTHROPLASTY, 1996, 11 (03) :351-353
[6]  
DAVIDSON JA, 1992, P 7 INT C BIOM ENG, P152
[7]  
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[8]  
Devane PA, 1995, CLIN ORTHOP RELAT R, V319, P303
[9]  
GRIFFITH MJ, 1978, CLIN ORTHOP RELAT R, V137, P37
[10]  
HARBAUGH ME, 1992, RECENT ADVANCES IN COMPUTER METHODS IN BIOMECHANICS & BIOMEDICAL ENGINEERING, P270