Influence of the errors made in the measurement of the extra-column volume on the accuracies of estimates of the column efficiency and the mass transfer kinetics parameters

被引:80
作者
Gritti, Fabrice
Felinger, Attila
Guiochon, Georges
机构
[1] Univ Tennessee, Dept Chem, Knoxville, TN 37996 USA
[2] Oak Ridge Natl Lab, Div Chem Sci, Oak Ridge, TN 37831 USA
[3] Univ Pecs, Dept Analyt Chem, H-7624 Pecs, Hungary
基金
美国国家科学基金会;
关键词
accuracy of HETP; column efficiency; HETP; van Deemter curves; extra-column volume; mass transfer; solute dispersion; sunfire-C-18; phenol;
D O I
10.1016/j.chroma.2006.09.074
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The influences of the errors made in the measurement of the extra-column volume of an instrument on the accuracies of the estimates made of the column efficiency and of the parameters of the mass transfer kinetics were investigated from an experimental point of view. A standard HP 1090 apparatus (extra-column volume, approximate to 50 mu L) was used to measure the efficiency of a Sunfire-C-18 RPLC column (column hold-up volume, -1.50 mL). The first and second moments of the peaks of phenol (a retained compound) and of thiourea (a practically non-retained compound) were measured at six different temperatures between 22 and 78 degrees C, for flow rates between 0.10 and 4.70 mL/min (i.e., for linear velocities between 0.025 and 1.179 cm/s). Each series of measurements was successively made with the instrument being fitted with and without the column. The experimental HETP data must be corrected for the solute dispersion in the connected tubes in order properly to assess the true column efficiency. Even with a modem, high performance instrument, the dispersion of a non-retained compound is essentially due to the band broadening phenomena that take place in the extra-column volumes, the sum of all these extra-column band broadening contributions accounting for more than 80% of the total band broadening measured. The contribution of the sampling device is particularly deleterious since, for a 2 mu L injection, the maximum solute concentration in the peak that enters into the column is nearly ten-fold lower than that of the sample. Nevertheless, the impact of the extra-column volumes on the estimates of the kinetic parameters (e.g., molecular diffusion coefficient D. and effective particle diffusivity De) remains negligible. Obviously, the relative error made on the column efficiency of a retained compound depends much on its retention factor. It decreases from 8 to 1% when the retention factor increases from 5 to 17. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:57 / 72
页数:16
相关论文
共 40 条
[1]   Determination of the porosities of monolithic columns by inverse size-exclusion chromatography [J].
Al-Bokari, M ;
Cherrak, D ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 2002, 975 (02) :275-284
[2]   ON THE DISPERSION OF A SOLUTE IN A FLUID FLOWING THROUGH A TUBE [J].
ARIS, R .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1956, 235 (1200) :67-77
[3]   DISPERSION OF PEAKS BY SHORT STRAIGHT OPEN TUBES IN LIQUID-CHROMATOGRAPHY SYSTEMS [J].
ATWOOD, JG ;
GOLAY, MJE .
JOURNAL OF CHROMATOGRAPHY, 1981, 218 (1-3) :97-122
[4]   A PRACTICAL METHOD TO PREDICT THE EFFECT OF EXTRA-COLUMN VARIANCE ON OBSERVED EFFICIENCY IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
COHEN, KA ;
STUART, JD .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1987, 25 (09) :381-390
[5]  
COQ B, 1981, J CHROMATOGR SCI, V19, P1
[6]   Repeatability and reproducibility of high-concentration data in reversed-phase liquid chromatography -: III.: Isotherm reproducibility on Kromasil C18 [J].
Felinger, A ;
Gritti, F ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1024 (1-2) :21-38
[7]  
Felinger A., 1998, DATA ANAL SIGNAL PRO
[8]   Adsorption isotherms of nonionic surfactants in SBA-15 measured by micro-column chromatography [J].
Findenegg, G. H. ;
Eltekov, A. Y. .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1150 (1-2) :236-240
[9]   INFLUENCE OF INSTRUMENTAL CONTRIBUTIONS ON APPARENT COLUMN EFFICIENCY IN HIGH-SPEED GAS-CHROMATOGRAPHY [J].
GASPAR, G ;
ANNINO, R ;
VIDALMADJAR, C ;
GUIOCHON, G .
ANALYTICAL CHEMISTRY, 1978, 50 (11) :1512-1518
[10]  
Giddings J.C., 1991, Unified separation science