Biomechanical evaluation of different fixation plates in medial opening upper tibial osteotomy

被引:13
作者
Esenkaya, Irfan [1 ]
Misirlioglu, Mesut
Kelestemur, M. Halidun
Elmali, Nurzat
Fadillioglu, Ersin
机构
[1] Inonu Univ, Sch Med, Dept Orthopaed & Traumatol, TR-44065 Malatya, Turkey
[2] Firat Univ, Sch Eng, Dept Met & Mat Engn, Elazig, Turkey
[3] Inonu Univ, Sch Med, Dept Physiol, TR-44065 Malatya, Turkey
关键词
knee; unicompartmental osteoarthritis; proximal tibial osteotomy; high tibial osteotomy; biomechanics; plate designs;
D O I
10.1016/j.knee.2006.10.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
In this biomechanical study, 25 in vitro calf tibial models were used in order to compare the stability of the plates under axial compression loading. A 10-mm medial opening gap was stabilized in each of the five calf tibial models either with four or two-holed rectangular shaped plates with wedges, with four-holed reversed L-shaped plates with wedges, with the combination of these two types of plates, or with six-holed anatomical T-plates. The compression behavior of the model was tested by using a universal mechanical testing system. The specimens fixed with the combination of plates with the four-holed reversed L-shaped and with two-holed rectangular shaped; or with six-holed anatomical T-plates, showed significantly better stability than those of others. Four different kinds of failure (slippage of wedge, lateral cortex fracture, damage and/or loosening of screws, and bending of plates) were observed on the models. When the average value of force loading on the plates that were designed by the first author was considered, the plates were stable and the average force values at these points were higher than the loading force on a knee during the normal paced walking or running conditions. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:46 / 50
页数:5
相关论文
共 24 条
[1]
Primary stability of four different implants for opening wedge high tibial osteotomy [J].
Agneskirchner, JD ;
Freiling, D ;
Hurschler, C ;
Lobenhoffer, P .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2006, 14 (03) :291-300
[2]
High tibial osteotomy with a calibrated osteotomy guide, rigid internal fixation, and early motion - Long-term follow-up [J].
Billings, A ;
Scott, DF ;
Camargo, MP ;
Hofmann, AA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (01) :70-79
[3]
A new distractor with angle-scale for proximal tibia medial opening wedge osteotomy [J].
Esenkaya, I .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2006, 14 (05) :443-446
[4]
Esenkaya Irfan, 2005, Acta Orthop Traumatol Turc, V39, P211
[5]
Proximal tibia medial open-wedge osteotomy using plates with wedges: early results in 58 cases [J].
Esenkaya, Irfan ;
Elmali, Nurzat .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2006, 14 (10) :955-961
[6]
FLAMME CH, 1994, ARCH ORTHOP TRAUMA S, V119, P450
[7]
Franco V, 2002, TECH KNEE SURG, V1, P43
[8]
Hanssen AD, 2001, SURG KNEE, P1447
[9]
Biomechanical superiority of plate fixation for proximal tibial osteotomy [J].
Hartford, JM ;
Hester, P ;
Watt, PM ;
Hamilton, D ;
Rohmiller, M ;
Pienkowski, D .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (412) :125-130
[10]