Distribution of inhaled nitric oxide during sequential and continuous administration into the inspiratory limb of the ventilator

被引:18
作者
Mourgeon, E [1 ]
Gallart, L [1 ]
Rao, GSU [1 ]
Lu, Q [1 ]
LawKoune, JD [1 ]
Puybasset, L [1 ]
Coriat, P [1 ]
Rouby, JJ [1 ]
机构
[1] HOP LA PITIE SALPETRIERE, DEPT ANESTHESIE, UNITE REANIMAT CHIRURG, F-75013 PARIS, FRANCE
关键词
nitric oxide; distribution; uptake; monitoring; ARDS;
D O I
10.1007/s001340050421
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: The concentrations of nitric oxide (NO) in the ventilatory circuits and the patient's airways were compared between sequential (SQA) and continuous (CTA) administration during inspiratory limb delivery. Design: Prospective controlled study. Setting: 14-bed Surgical Intensive Care Unit of a teaching University hospital. Patients and participants: Eleven patients with acute lung injury on mechanical ventilation and two healthy volunteers. Interventions: A prototype NO delivery device (Opti-NO) and Cesar ventilator were set up in order to deliver 1, 3 and 6 parts per million (ppm) of NO into the bellows of a lung model in SQA and CTA. Using identical ventilatory and Opti-NO settings, NO was administered to the patients with acute lung injury. Measurements and results: NO concentrations measured from the inspiratory limb [INSP-NOMeas] and the trachea [TRACH-N-Meas] using fast response chemiluminescence were compared between the lung model and the patients using controlled mechanical ventilation with a constant inspiratory flow. INSP-NOMeas were stable during SQA and fluctuated widely during CTA (fluctuation at 6 ppm = 61 % in the lung model and 58 +/- 3 % in patients). Inpatients, [TRACH-NOMeas] fluctuated widely during both modes (fluctuation at 6 ppm = 55 +/- 3 % during SQA and 54 +/- 5 % during CTA). The NO now requirement was significantly lower during SQA than during CTA (74 +/- 0.5 vs 158 +/- 2.2 ml.min(-1) to attain 6 ppm, p = 0.0001). INSP-NOMeas were close to the Values predicted using a classical formula only during SQA (bias = -0.1 ppm, precision = +/- 1 ppm dur ing SQA; bias = 2.93 ppm and precision = +/- 3.54 ppm during CTA). During SQA, INSP-NOMeas varied widely in healthy volunteers on pressure support ventilation. Conclusions: CTA did not provide homogenous mixing of NO with the tidal volume and resulted in fluctuating INSP-NOMeas. In contrast, SQA delivered stable and predictable NO concentrations during controlled mechanical ventilation with a constant inspiratory flow and was economical compared to CTA. However, SQA did not provide stable and predictable NO concentrations during pressure support ventilation.
引用
收藏
页码:849 / 858
页数:10
相关论文
共 18 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Alveolar neutrophil functions and cytokine levels in patients with the adult respiratory distress syndrome during nitric oxide inhalation [J].
CholletMartin, S ;
Gatecel, C ;
Kermarrec, N ;
GougerotPocidalo, MA ;
Payen, DM .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1996, 153 (03) :985-990
[3]   A study of mixing conditions during nitric oxide administration using simultaneous fast response chemiluminescence and capnography [J].
Foubert, L ;
Mareels, K ;
Fredholm, M ;
Lundin, S ;
Stenqvist, O .
BRITISH JOURNAL OF ANAESTHESIA, 1997, 78 (04) :436-438
[4]  
GROVER ER, 1994, ANAESTH INTENS CARE, V22, P312
[5]   Inaccuracies of nitric oxide delivery systems during adult mechanical ventilation [J].
Imanaka, H ;
Hess, D ;
Kirmse, M ;
Bigatello, LM ;
Kacmarek, RM ;
Steudel, W ;
Hurford, WE .
ANESTHESIOLOGY, 1997, 86 (03) :676-688
[6]   ABSORPTION OF NITROGEN-DIOXIDE AND NITRIC-OXIDE BY SODA LIME [J].
ISHIBE, T ;
SATO, T ;
HAYASHI, T ;
KATO, N ;
HATA, T .
BRITISH JOURNAL OF ANAESTHESIA, 1995, 75 (03) :330-333
[7]   A NEW SIMPLIFIED ACUTE PHYSIOLOGY SCORE (SAPS-II) BASED ON A EUROPEAN NORTH-AMERICAN MULTICENTER STUDY [J].
LEGALL, JR ;
LEMESHOW, S ;
SAULNIER, F .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1993, 270 (24) :2957-2963
[8]   DOSE-RESPONSE CARVES OF INHALED NITRIC-OXIDE WITH AND WITHOUT INTRAVENOUS ALMITRINE IN NITRIC OXIDE-RESPONDING PATIENTS WITH ACUTE RESPIRATORY-DISTRESS SYNDROME [J].
LU, Q ;
MOURGEON, E ;
LAWKOUNE, JD ;
ROCHE, S ;
VEZINET, C ;
ABDENNOUR, L ;
VICAUT, E ;
PUYBASSET, L ;
DIABY, M ;
CORIAT, P ;
ROUBY, JJ .
ANESTHESIOLOGY, 1995, 83 (05) :929-943
[9]  
MOORS AH, 1994, ANAESTH INTENS CARE, V22, P310
[10]   AN EXPANDED DEFINITION OF THE ADULT RESPIRATORY-DISTRESS SYNDROME [J].
MURRAY, JF ;
MATTHAY, MA ;
LUCE, JM ;
FLICK, MR .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 138 (03) :720-723