Comparison of the accuracy of the lithium dilution technique with the thermodilution technique for measurement of cardiac output

被引:94
作者
Kurita, T
Morita, K
Kato, S
Kikura, M
Horie, M
Ikeda, K
机构
[1] Dept. Anaesthiol. and Intensive Care, Hamamatsu Univ. School of Medicine, Hamamatsu, 431-31
关键词
heart; cardiac output; measurement techniques; thermodilution; lithium dilution; flowmetry; pig;
D O I
10.1093/bja/79.6.770
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A new indicator dilution technique for measurement of cardiac output is described. Lithium chloride is injected via a central venous catheter and its dilution curve measured in arterial blood using a lithium-selective electrode. We assessed the lithium dilution cardiac output measurement (LiDCO) and a conventional thermodilution cardiac output measurement (ThDCO) by comparing the results of both with cardiac output determined by electromagnetic flowmetry (EMCO) under controlled laboratory conditions in 10 swine. They were monitored with a pulmonary artery catheter, femoral artery catheter and electromagnetic flowmeter placed around the ascending aorta. LiDCO, ThDCO and EMCO measurements were determined at baseline, in a hyperdynamic state produced by administration of dobutamine, at a second baseline and finally in a hypodynamic state induced by propranolol during deep anaesthesia. Data were analysed by linear regression analysis and the comparison method described by Bland and Altman; bias and precision of both LiDCO and ThDCO compared with EMCO were calculated by the method of Sheiner and Beal. The correlation coefficient between LiDCO and EMCO (0.95) was higher than that between ThDCO and EMCO (0.87). The precision value of LiDCO (0.04) was significantly less (i.e. better) than that of ThDCO (0.09). The results of this study indicated that LiDCO was more reliable compared with conventional ThDCO.
引用
收藏
页码:770 / 775
页数:6
相关论文
共 35 条
[1]   EFFECTS OF THE RESPIRATORY CYCLE ON CARDIAC-OUTPUT MEASUREMENTS - REPRODUCIBILITY OF DATA ENHANCED BY TIMING THE THERMODILUTION INJECTIONS IN DOGS [J].
ARMENGOL, J ;
MAN, GCW ;
BALSYS, AJ ;
WELLS, AL .
CRITICAL CARE MEDICINE, 1981, 9 (12) :852-854
[2]   ALTERATION OF PULMONARY BLOOD-FLOW BY PULMONARY-ARTERY OCCLUDED PRESSURE MEASUREMENT [J].
BERRY, AJ ;
GEER, RT ;
MARSHALL, BE .
ANESTHESIOLOGY, 1979, 51 (02) :164-166
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   The effectiveness of right heart catheterization in the initial care of critically ill patients [J].
Connors, AF ;
Speroff, T ;
Dawson, NV ;
Thomas, C ;
Harrell, FE ;
Wagner, D ;
Desbiens, N ;
Goldman, L ;
Wu, AW ;
Califf, RM ;
Fulkerson, WJ ;
Vidaillet, H ;
Broste, S ;
Bellamy, P ;
Lynn, J ;
Knaus, WA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11) :889-897
[5]   A PROSPECTIVE ANALYSIS OF 1400 PULMONARY-ARTERY CATHETERIZATIONS IN PATIENTS UNDERGOING CARDIAC-SURGERY [J].
DAMEN, J ;
BOLTON, D .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1986, 30 (05) :386-392
[6]   RENAL EXCRETION OF CATION IN THE DOG DURING INFUSION OF ISOTONIC SOLUTIONS OF LITHIUM CHLORIDE [J].
FOULKS, J ;
MUDGE, GH ;
GILMAN, A .
AMERICAN JOURNAL OF PHYSIOLOGY, 1952, 168 (03) :642-649
[7]   NEW TECHNIQUE FOR MEASUREMENT OF CARDIAC OUTPUT BY THERMODILUTION IN MAN [J].
GANZ, W ;
DONOSO, R ;
MARCUS, HS ;
FORRESTER, JS ;
SWAN, HJC .
AMERICAN JOURNAL OF CARDIOLOGY, 1971, 27 (04) :392-+
[8]  
GROBAN L, 1993, ANESTHESIOLOGY, V79, P1444
[9]   THE SLOWING OF SINUS RHYTHM DURING THERMODILUTION CARDIAC-OUTPUT DETERMINATION AND THE EFFECT OF ALTERING INJECTATE TEMPERATURE [J].
HARRIS, AP ;
MILLER, CF ;
BEATTIE, C ;
ROSENFELD, GI ;
ROGERS, MC .
ANESTHESIOLOGY, 1985, 63 (05) :540-541
[10]   VENTRICULAR-FIBRILLATION DURING THERMODILUTION CARDIAC-OUTPUT DETERMINATION [J].
KATZ, RI ;
TELLER, ED ;
POPPERS, PJ .
ANESTHESIOLOGY, 1985, 62 (03) :376-377